Sterilization cycles
Prepare the sterilizer
WARNINGS
Notice: Use only distilled or demineralized water (see "Feed water
specifications (EN 13060)" on page 100 for technical
requirements). Do not add any chemical / additive to the water.
HOOKING AND REMOVING THE STEAM DIFFUSER PLATE
WARNING! Ensure that the steam diffuser plate is firmly
hooked in its position before starting a sterilization cycle.
An improper positioning of the steam diffuser plate could
result in bad steam quality and could impair the
sterilization process, with risk of non sterile load and cross
infection.
Sterility at the end of the cycle is not guaranteed if the
steam diffuser plate was not correctly placed.
To hook the steam diffuser plate, slide it into the
chamber until il gets engaged into the end hooks.
To remove the steam diffuser plate, press it in
the center of the end edge (1) and slide it
outwards (2).
FILLING THE CLEAN WATER TANK
1
Switch the sterilizer ON and remove the tank filling cover-cap.
2
Fill the clean water tank with distilled or demineralized water
until the sterilizer makes a sound. See "Technical data" on
page 96 for the tank volume.
3
Reposition the tank filling cover-cap.
INSERTING THE CHAMBER RACK INTO THE STERILIZER
CAUTION! Risk of burns. Before touching the chamber rack
or contents, ensure the sterilization chamber is not hot.
1
Open the chamber door and align the chamber rack at the
center/bottom of the chamber.
2
Push the chamber rack gently into position until it clicks
into place.
3
Insert cassettes horizontally or vertically, or insert trays.
See "Load maintenance and preparation" on page 43 for
load requirements and "Chamber accessories" on page 17.
4
Close the door.
5
Turn the sterilizer switch ON: after the initialization the
homepage appears.
GENERAL RECOMMENDATIONS
Follow these recommendations to obtain the most from the drying:
n
Ensure the paper side of the sterilization bags faces up, and
that the space in-between bags is enough.