Operating the analyzer 

40 

Waltron User Manual 101-045-E.1 
9031CX Sodium Analyzer 

Average factor: increasing the average factor will increase noise dampening and decrease 
response time of the analyzer.  Recommended Average factor = 80 
Sensor data: this is a free text field up to 12 characters. The user may input the date of 
exchange of a new sensor, the name of a sensor, etc. 
Backlight Off delay:  user may adjust the time the analyzer takes to turn off the backlight on 
the display in order to save the life of the HMI screen / save energy.  This may be from 0 min 
to 120 mins. 

 

3.6 

TROUBLESHOOTING  

3.6.1 

CALIBRATION FAIL 

A Calibration Fail condition will occur after a 2-Pt calibration if the electrode response does not 
meet requirements.  This happens when the electrodes “strength” (“slope”) from a 2-Pt calibration 
is below 83 or greater than 110. This could be caused by a number of factors 

3.6.1.1 

Strength less than 83% 

 

Make sure standard containers (CAL1 and CAL2) are full of solution. 

 

Check that vapor bubbles are emerging from the bottom of the stainless steel entrainment tube

.

 

 

Check the condition of the reagent solution. 

 

Regenerate the sodium measuring electrode.  See Section 

Monthly maintenance

 

If the slope value is not improved after the sodium measuring electrode is regenerated, the electrode 
may need to be replaced. 

3.6.1.2 

Strength more than 100% 

 

Possible cause may be faulty reference electrode.  Check status of reference electrode filling solution 
and add more if using N3010-171A. 

 

Check the status of reference electrode and replace if necessary. 

 

Make sure standard containers (CAL1 and CAL2) are full of solution. 

3.6.1.3 

Very low or 0% slope 

 

Check the operation of the solenoid valve(s). 

 

Check flow of standard solution through flow cell. 

 

Check the level of the salt bridge solution in the reference electrode, if using N3010-171A. 

 

Check for open circuit reference electrode by substituting it with an electrode of known performance. 

 

Check all electrical connections in the electrode junction box and interconnect cable. 

 

Make sure standard containers (CAL1 and CAL2) are full of solution. 

 

Make sure the low 100ppb standard is connected to Cal1 tubing and high 1000ppb standard is 
connected to Cal2 tubing.  Note that the high standard is red in color. 

 
 
 
 
 
 

Содержание 9031CX

Страница 1: ...101 045 E 1 EXPERTS IN WATER CHEMISTRY SINCE 1903 9031CX Sodium Analyzer User Manual...

Страница 2: ...46 www waltron net Please be ready to provide the following information Date analyzer was purchased Analyzer model and serial number Recent maintenance history Calibration slope values and detailed de...

Страница 3: ...s containers packages and chemicals Only qualified personnel should be involved in the installation operation and servicing of the analyzer Follow safety precautions when operating analyzer in conditi...

Страница 4: ...cal Service Department Customer name address and department Name and telephone number of the individual responsible for returning items for repair Brief problem description Ship to Waltron service cen...

Страница 5: ...ng and shipping needs Check all items received against those on the packing list Chemicals are usually shipped in a separate package and will be itemized accordingly Verify that the number of packages...

Страница 6: ...which if not avoided may result in minor or moderate injury It may also be used to alert against unsafe practices CAUTION symbol on the equipment refers the user to the product manual for additional...

Страница 7: ...sted and meets applicable standards for safety and or performance The Ex mark means the equipment complies with the requirements of the European standards that are harmonised with the 94 9 EC Directiv...

Страница 8: ...it 13 2 1 3 Sample requirements 14 2 1 4 External piping connections 15 2 2 Electrode installation 15 2 2 1 Probe connections 15 2 2 2 Probe installation 16 2 2 3 Automatic KCl Refill System optional...

Страница 9: ...nance 41 4 1 Buffer solution s 41 4 1 1 Concentrated Ammonia Solution 1 liter 41 4 1 2 Diisopropylamine Solution 99 41 4 2 Standard solutions 42 4 2 1 Preparing standard solutions 42 4 3 Etching solut...

Страница 10: ...kage and will be itemized accordingly Verify that the number of packages received agrees with the packing list and shipping papers Notify both Waltron and the carrier if any problems occur 1 1 2 MAIN...

Страница 11: ...e following 1 Wet Section Sensor Unit a Hydraulic panel consisting of constant head flowcell tubing thermistor and solenoid valve b Sodium and reference electrodes c Reagent container d Separate sodiu...

Страница 12: ...al system and converted into a ppb ppm measurement An internal thermistor housed in flowcell is used to monitor the sample temperature for temperature compensation If the sample temperature exceeds 13...

Страница 13: ...1 LOCATION AND LAYOUT Mount the analyzer in a clean vibration free area avoiding direct radiant heat sunlight and drafts Avoid areas containing chlorinating equipment The 9031CX Sodium analyzer defau...

Страница 14: ...trols the operations of the analyzer Power supply CPU card DIO card and input power terminal junction are housed in the transmitter enclosure Transmitter unit is a NEMA 4X rated enclosure with cable g...

Страница 15: ...ions for the transmitter mounting bracket 2 1 3 SAMPLE REQUIREMENTS The maximum sample pressures and temperatures specified in the SPECIFICATION section should not be exceeded The sample should be int...

Страница 16: ...f the wet section case with sample inlet and drain connections 2 2 ELECTRODE INSTALLATION 2 2 1 PROBE CONNECTIONS Figure 7 Probe connections Sodium Measuring Electrode Reference Electrode Refillable v...

Страница 17: ...gh the clamp and into the flowcell Electrodes may have to be inserted at an angle 4 For reference electrode insert the electrode so that the tip of the electrode is 5mm into the sample Correct placeme...

Страница 18: ...ubing through nozzle of bottle and connect outer tubing over the nozzle of bottle Make sure the inner tubing reaches the top of the bottle while the other end remains inside the reference electrode 5...

Страница 19: ...Sodium Analyzer 2 3 ELECTRICAL CONNECTIONS 2 3 1 WET SECTION UNIT Figure 9 Wet section layout and connections Pre Amp Reagent Buffer Sodium Measuring Electrode Reference Electrode Thermistor Constant...

Страница 20: ...E AMP UNIT The pre amplifier unit has 3 connection terminals J1 J2 J3 shown below Figure 10 Pre amplifier layout Pin 1 of all the connectors is shown in black Pre Amp Output to Transmitter J3 Connect...

Страница 21: ...J3 Pin 1 Black Wire Pin 2 White Black Wire Pin 3 Red Wire Pin 4 White Wire Pin 5 Red Black Wire Pin 6 Green Wire Pin 7 Blue Wire Pin 8 Orange Wire Pin 4 Shield Optional 5 Multicore cable 2 Core PVC Sh...

Страница 22: ...nsmitter case ATTENTION Before connecting the analyzer to the main power supply check that there is correct voltage at the mains WARNING Although this instrument has internal fuse protection the opera...

Страница 23: ...is connected to the CPU Card J4 as follows Pin 1 Black Wire Pin 2 White Black Wire Pin 3 Red Wire Pin 4 White Wire Pin 5 Red Black Wire Pin 6 Green Wire Pin 7 Blue Wire Pin 8 Orange Wire 2 3 5 CONNEC...

Страница 24: ...M 6 o Pin 6 NO 6 High Alarm o Pin 7 NC 7 o Pin 8 COM 7 o Pin 9 NO 7 2 3 7 CONNECTINGTHESOLENOIDTOTRANSMITTER The solenoid valveusedforcalibrationisterminatedontheCPUcardJ5 The 24V DC excitation 400mA...

Страница 25: ...Operating the analyzer 24 Waltron User Manual 101 045 E 1 9031CX Sodium Analyzer 3 OPERATING THE ANALYZER 3 1 ANALYZER OPERATION Figure 128 Sample flow during normal operation...

Страница 26: ...solution is sent to the flowcell where it comes into contact with the sodium and reference electrodes The sample flows past the electrodes and exits the flowcell through the drain located in the bott...

Страница 27: ...ed at 2A 250VAC non inductive See Figure 10 for location of alarm relays Concentration alarm descriptions Symbol ALARM DESCRIPTION CONC LOW Low Alarm Activates when Na in sample feed is lower then Low...

Страница 28: ...cket terminal connector located in transmitter section and switch ON the system 2 The analyzer automatically displays the concentration of sodium interpreted by the electrodes 3 If the analyzer is bei...

Страница 29: ...rectly Before initiating a calibration sequence rinse the two standard solution containers with high purity DI water and fill them with fresh sodium standard solutions This establishes the slope of th...

Страница 30: ...er must then connect the low 100ppb standard solution and press the START button when complete After the START button is pressed the analyzer will begin counting down the remaining time until the sole...

Страница 31: ...flowcell The analyzer calibrates by changing the offset of the strength slope taken during the last 2 Pt calibration To perform Process Calibration From the Commands Window press the PROCESS CAL butt...

Страница 32: ...ORT OPERATION Any calibration cycle or Grab Sample cycle may be interrupted at any time by pressing the ABORT OPERATION button Once the calibration process is aborted the measurement screen is display...

Страница 33: ...sensor reading in mV Each vertical line represents two minutes The Y axis represents the sensor output in mV To access the Display Window Sensor Potential reading Temperature Temperature reading of t...

Страница 34: ...three alarm relays and the 4 20mA output To access the Checks Window From the Main Menu press the DISPLAY button Then press the CHECKS button The buttons for the valves and the relays will toggle on...

Страница 35: ...pped with internal data loggers Each data logger will automatically record your data and will automatically overwrite the oldest data set and record over with the most recent data set when the log rea...

Страница 36: ...p or down to view results for the current day selected and select which day to view by selecting the day from the drop down menu box in the upper right corner Dates are displayed in YY MM DD format Re...

Страница 37: ...3 CALIBRATIONS DATALOG Results of the last 10 calibrations are stored in the Calibration Datalog Each result is recorded with the Time and Date the calibration occurred Strength or slope Outcome pass...

Страница 38: ...password protected To access Service Mode From the Main Menu select service from the drop down menu at the top of the screen After service has been selected then press on the to open the keypad Enter...

Страница 39: ...ld be set to 100 ppb Standard sol 2 this should represent the concentration of the high sodium standard and should be set to 1000ppb 3 5 10 GENERAL SETTINGS WINDOW The General Settings Menu will allow...

Страница 40: ...2W Modbus param Slave ID adjust for Slave identification number BaudRate 9600 DataBits 8 Parity Even StopBit 1 Valves Timing min Variables in this block set the timing for the calibration sequences D...

Страница 41: ...e bottom of the stainless steel entrainment tube Check the condition of the reagent solution Regenerate the sodium measuring electrode See Section Monthly maintenance If the slope value is not improve...

Страница 42: ...tainer of ammonia and the cap must be released with care Waltron uses a 28 30 w v solution 0 89 s g which is recommended Refer to section 5 7 which relates to sodium concentration and pH ATTENTION Wal...

Страница 43: ...ric flask Make up to the 1 liter mark with high purity water to give a solution of 10ppm sodium ions Pipette 20ml of the 10ppm solution into a two liter volumetric flask and make up to the 2 liter mar...

Страница 44: ...rade sodium fluoride NaF in approximately 400ml high purity water Add to this solution 20 0 2 ml 5M acetic acid CH3COOH and dilute to 1 liter ATTENTION 5M acetic acid can be prepared from concentrated...

Страница 45: ...Regeneration Instructions When the sodium concentration is below 1ppb apply the following regeneration etch procedure before carrying out a Two Point Calibration ATTENTION When used for prolonged per...

Страница 46: ...iner and safely dispose of the solution Rinse the containers thoroughly WARNING For safe handling instructions of buffer solutions refer to Section 4 Fill the Low Sodium Standard 100ppb calibration so...

Страница 47: ...ation but the pH value is calculated as pH must be greater than pNa 3 so ideally at 100ppb Na the pH must be greater than 8 4 10ppb Na the pH must be greater than 9 4 1ppb Na the pH must be greater th...

Страница 48: ...Di Isopropylamine 1 Gallon 5 2 RECOMMENDED SPARE PARTS PART NUMBER DESCRIPTION N3010 177 Electrode Measuring N3010 174 Electrode Reference Re fillable N3010 173 Electrode Reference Gel filled N3010 1...

Страница 49: ...m into the sample Calibration Fail Strength is too low 83 Ensure calibration standard is flowing Perform Solenoid Check cycle in Diagnostics to verify Calibration Fail Strength is too low 83 Regenerat...

Страница 50: ...tion log and run another calibration if results are not good Readings are not accurate too high Poor electrode performance Bad calibration Sodium leak in sample system Regenerate replace sodium measur...

Страница 51: ...tep 5 minutes Current Outputs Two isolated 4 20mA current analog outputs Alarms Three voltage free contacts alarm points set from transmitter rated at 2A 250VAC non inductive Power Wide range of input...

Страница 52: ...5 ppb 435 27 0 1 ppb 417 47 0 2 ppb 399 66 0 5 ppb 377 11 1 ppb 358 31 2 ppb 340 5 5 ppb 316 95 10 ppb 299 15 20 ppb 281 34 50 ppb 257 79 100 ppb 239 99 200 ppb 222 18 500 ppb 198 63 1 ppm 180 83 2 pp...

Страница 53: ...ENDIX PANEL MOUNT This Appendix shows the schematics of the Retrofit Panel Mount Bracket option for the transmitter case for panel mounting over an existing 9030 panel cut out and the dimensions of th...