Operating the analyzer
26
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
3.2
ALARMS
3.2.1
SAMPLE CONCENTRATION ALARMS
When the 9031CX Sodium system is in normal operation mode one alarm operates as a “low”
alarm and the other operates as a “high” alarm. The CONC LOW alarm is activated when the
sodium level decreases below the set value. The CONC HIGH alarm is activated when the
sodium level increases above the set value. The two sodium alarms control the relays
provided. Each relay has one pair of changeover contacts rated at 2A, 250VAC (non-
inductive). S ee F igure 10 f or location of alar m r elays
Concentration alarm descriptions
Symbol
ALARM
DESCRIPTION
CONC LOW
Low Alarm
Activates when Na in sample feed is
lower then “Low Set Point”.
CONC HIGH
High Alarm
Activates when Na in sample feed is
higher then “High Set Point”.
3.2.2
ALARM DESCRIPTIONS
ALARM
DESCRIPTION
CALIBRATION FAIL
Unit failed calibration
SAMPLE HOT
Sample temperature over range (131F)
T. SENSOR FAIL
No thermistor response
O/P 1 Out
Concentration is outside O/PmA 1 set range
O/P 2 Out
Concentration is outside O/PmA 2 set range
CONC LOW
Concentration is below Low Alarm set point
CONC HIGH
Concentration is above High Alarm set point
OVR
Concentration is above limits of analyzer (>10ppm)