Operating the analyzer
38
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
3.5.9
CALIBRATION SETUP WINDOW
The Calibration Set Up menu will allow you to make adjustments to the calibration settings. You may
adjust the concentration levels of the calibration standard.
To access the Calibration set up window:
From the Main Menu under service mode, press the “CONFIGURATION” button. Then press the
“CALIBRATION SET UP” button.
Standard sol.1:
this should represent the concentration of the low sodium standard and should be set to
100 ppb
Standard sol.2:
this should represent the concentration of the high sodium standard and should be set to
1000ppb
3.5.10
GENERAL SETTINGS WINDOW
The General Settings Menu will allow the user to set alarm ranges and time and date.
To access the General settings window:
From the Main Menu under service mode, press the “CONFIGURATION” button. Then press the
“GENERAL SETTINGS” button.
Low conc.alarm:
set the alarm limit for low concentration of sodium sample
High conc.alarm:
set the alarm limit for high concentration of sodium sample
Measure range:
set the measurement range
Temp.range:
enter the upper limit of your temperature range
Time & date:
enter the time and date for the analyzer’s internal clock