Gebrauchsanweisung Untersuchungseinheit cubeONE,
directions for use, refraction unit cubeONE
38
Wagner & Guder Medical GmbH
issue September 2020
Malfunction
Possible cause
Remedy
Link
Installed
230VAC device
without working
Look, no working
-
-
Fuse on circuit board
defective
Contact distributor
-
Fuse of device defective
Change fuse for device
Operating manual of
devices
Device on
instrument table-
no dimming
working
Potentiometer defective
Contact distributor
page 20
Cable break
Contact distributor
-
Disconnection to circuit
board
Contact distributor
page 25
Reading lamp do
not work
Power supply or
illuminants defective
Contact distributor
-
Power output reading
lamp defective
Contact distributor
page 25
Safety switch not
working
switch defective
Contact distributor
page 17
Safety switch damaged
Schaltkontakt wieder
herstellen
page 17
Cable break
Contact distributor
page 17
Disconnection to circuit
board
Contact distributor
page 25
2nd working
position not
accessilbe
2nd working position
fixed with screw
Disassemble screw
page 11
Cable guide blocked
moving
Contact distributor
-
Dirt in rail system
Clean sliding system
-