Gebrauchsanweisung Untersuchungseinheit cubeONE,
directions for use, refraction unit cubeONE
7
Wagner & Guder Medical GmbH
issue September 2020
Informations for the operator
Before using the device, make yourself thoroughly familiar with the
contents of the operating instructions. Also observe the operating
instructions for accessories and other system components.
The examination unit may only be installed and commissioned by
qualified personnel who are familiar with the installation,
commissioning and operation of the product. For the purposes of these
operating instructions, qualified personnel are persons who, on the
basis of their technical training, knowledge and experience and their
knowledge of the relevant standards, are able to assess the work
assigned to them and recognise possible risks.
Keep the operating instructions ready to hand at all times for the
operating and service personnel.
Observe the legal regulations for accident prevention and work safety
valid in the respective country.
Changes and repairs to the examination unit may only be carried out
by our service personnel or by designated authorised dealers. The
manufacturer is not liable for possible damage caused by unauthorised
interventions on the unit. In this case, all warranty claims are void.
The examination unit must not be installed and operated in damp
rooms. Avoid dripping and splashing water.
The examination unit must not be operated in a potentially explosive
environment.
The examination unit was adjusted to a level when it was installed. If
the unit has to be moved, please make sure that all adjusting elements
of the base plate touch the floor. This is the only way to ensure the
stability of the unit and the examination equipment.
In order to guarantee the specified working position of the examination
equipment, it is necessary to swivel and move the telescopic table and
the phoropter arm. Please make sure that the patient does not come
into contact with moving parts. There is a risk of injury!
The maximum load capacity of the instrument table is 40kg, while
instrument position 1 may be loaded with a maximum of 30kg.
Pay attention to the maximum load capacity of the patient chair
of 135 kg.
The lifting column of the patient chair is not designed for permanent
operation. After an operating time of more than 1 minute, the lifting
column of the patient chair requires a cooling time of 9 minutes.