Gebrauchsanweisung Untersuchungseinheit cubeONE,
directions for use, refraction unit cubeONE
37
Wagner & Guder Medical GmbH
issue September 2020
Technical error in the system
Solve technical errors
If an error occurs, shut down the system and use the following error list
to correct the error. If this does not remedy the error, mark the device as
non-functional and inform the person who put your entire system into
operation.
For your safety
Malfunction
Possible cause
Remedy
Link
No function
Power cord not
connected
Check connection
between power outlet
and refraction unit
page 13/ 18
Power failure
Notify home electrican
-
Power supply defective
Contact distributor
page 24
Main fuse defective
Change mains fuse
page 18
Devices on
instrument table
without working
Installed instruments
disconnected or disabled
Enable device or check
connection
Operating manual of
devices
Instrument table not in
working position
Move table in end
position
page 14
End switch defective
Contact distributor
page 21
Cable connection to the
examination device
disconnected (cable
break)
Contact distributor
page 20
Power supply defective
Contact distributor
page 24
Patient chair
without
working
Look, no working
-
-
Cable break control panel
Contact distributor
page 19
Electric pillar defective
Contact distributor
-
Safety switch activated or
blocked
Check safety switch and
control correct working
page 17