1.3.13 Limit of the permissible value of the analyzer readout setting time
t
у
in
analyte temperature measurement, min ………………………………………….….. 3.
1.3.14 Analyzer oscillatory readings in DOC measurement for 8 h, mg/dm
3
,
max …………………………………………………..……………...….. ±(0,025+0,02
C
).
1.4 Product components
The analyzer comprises:
converter unit;
oxygen sensor with a 5 m connecting cable;
protective cap.
1.5 Description and operation
1.5.1 General analyzer data
The MAPK-302E dissolved oxygen analyzer is a small-size microprocessor
device designed to measure the mass concentration of water-dissolved oxygen
(DOC) and the analyte temperature.
The DOC values measured in mg/dm
3
or temperature measured in Celsius
degrees (depending on the measuring mode) are displayed on the readout device
– a digital LCD display (“the display”). The minimum value of the least significant
digit for DOC measurement is 0,001 mg/dm
3
. The least significant digit value for
temperature measurement is 0,1 °С.
The analyzer is calibrated by atmospheric air of 100 % humidity with auto-
matic adjustment for atmospheric pressure during calibration.
Adjustment for atmospheric pressure during the analyzer calibration by at-
mospheric air is provided by a built-in atmospheric pressure sensor.
Note
– The DOC values may be displayed in % O
2
(in % saturation) with the
least significant digit value of 0,1 % O
2
.