1 DESCRIPTION AND OPERATION
1.1 Purpose
1.1.1 Product name and designation
MAPK-302E dissolved oxygen analyzer
TU 4215-022-39232169-2008
.
1.1.2 The analyzer has been designed to measure the dissolved oxygen
concentration and water temperature.
1.1.3 The analyzer may be used to measure the mass concentration of wa-
ter-dissolved oxygen and temperature of surface, waste and drinking water, and
may be employed by fish farms; chemical, biotechnological and food industries;
educational institutions and ecology branches.
1.1.4 The type of analyzer:
amperometric;
with external poling voltage;
with one sensitive element;
with a digital LCD display;
with automatic temperature compensation;
with dip sensor;
with automatic calibration when the sensor is placed in the oxygen envi-
ronment (air) at temperatures varying b15 °С and +35 °С;
with automatic adjustment for atmospheric pressure during calibration.
1.2 Basic parameters
1.2.1 By resistance to climatic effects, the analyzer version group is B4 in
accordance with GOST 12997-84.
1.2.2 By resistance to mechanical effects, the analyzer version is L1 in ac-
cordance with GOST 12997-84.
1.2.3 The converter unit degree of protection ensured by the casing is IP30
in accordance with GOST 14254-96.
1.2.4 By resistance to atmospheric pressure, the analyzer version group is
Р1 in accordance with GOST 12997-84 (atmospheric pressure between 84 and
106,7 kPa).
1.2.5 Analyzed water parameters:
temperature, °С ………………………………………………….…... 0 to +50;
pressure, MPa, max ……...………………………………..….……....… 0,05;
salt content, g/dm
3
……...………………………..………..……..…... 0 to 40;