18
Fig. 2.2
Turn the meter on. With button
“MODE”
choose indication of DOC in
mg/dm
3
or % sat.
The speed of water flow of analyte water along the membrane should be min
5 cm/s. In stationary fluid the measurements would lower slowly.
The measuring process before getting the fixed value is max 2 min.
Notes
1 If the temperature of the probe and aquatic environment differs, the
readout setting time is 3-5 min.
2 If there is no stir bar, it is recommended to submerge in the flask glass or
plastic balls and stir arm circling the flask till the readout set (about 2 min).
If values of dissolved oxygen decrease slowly and do not set in admitted
time, then there may be air bubbles on the membrane. If so, it is important:
while measuring in reservoir, stir the probe harshly leaving it submerged in
the water;
while measuring in BOD flask, invert the flask with the probe in it and stir.
In viscous medium it is necessary to clean the membrane from time to time
(para.2.6.2).
2.4.4
Water temperature measurement
The temperature measurement mode
t °С
is enabled by pressing the
MODE
button.
Probe
Rubber ring
Flask
Stir bar