26
2.6.6.1 Zero shift setting
1
Switch off the analyzer.
2
Press the
CAL
button and, keeping it pressed, switch on the analyzer. Af-
ter a sound signal release the
CAL
button. The screen will display
с2
.
3
Press the
ENTER
button. The
donЕ
capture will briefly appear on the
screen and the analyzer will go into the measurement mode. The display will indi-
cate readings in mg/dm
3
with zero shift.
2.6.6.2 Average slope setting
1
Switch off the analyzer.
2
Press the
CAL
button and keeping it pressed, switch on the analyzer. Af-
ter a sound signal release the
CAL
button. The screen will display
с2
. Press the
CAL
button again. The screen will display
с3
.
3
Press the
ENTER
button. The
donЕ
capture will briefly appear on the
screen and the analyzer will go into the measurement mode. The display will indi-
cate readings in mg/dm
3
corresponding to the average sensor slope.
Once the analyzer initial parameters are set, proceed to para.2.3.4.