5
vwr.com
I Instruction manual VWR
®
Media Pump
1
Package Contents
Description
Quantity
Peristaltic Watson Marlow
TM
pump head
1
Set of 3.2 mm dispensing assembly ( GL 45 bottle
con tube weight)
1
Set of power cables (EU, UK, CH)
1
User’s manual
1
Conformity certificate
1
Installation
To use
Media Pump
at its best, we suggest installing it on a stable, flat and rigid surface.
Example with bottle container:
Example with broth bag:
Put the tube into the peristaltic pump (from front to back). Be careful to center it to avoid pinching the tubing
when closing the pump. (BE CAREFUL: carefully read the tubing installation guide provided).
When the pump is opened, you must adjust it according to the inner
diameter of the tubing by moving the rotating button.
Turn the wheel on each side of the pump to get the correct setting
according to the diameter (1.6 / 3.2 / 4.8 / 6.4 / 8 mm) and the
thickness of the tubing (1.6 mm).
Insert the tubing into the pump.
Keep the tubing stretched while you close the pump.
WRONG
RIGHT
The tubing is pinched.
It can be pierced.
The tubing is not pinched.
There is no risk.
2
3
4