VPFlowScope M probe
30
© 2016 Van Putten Instruments BV | MAN-VP-SMPR-UK | Revision:1600 | Date:23-03-2016
3.3 Boundaries
Set upper and/or lower boundaries for alarm triggers
3.4 Period
Set the hold and reset period for alarm events
3.5 Reset
Reset the alarm counter. The alarm counter can only be reset to zero.
4 Info/about
This option shows all general information of the transmitter and the VPSensorCartridge, if connected.
5 Reset
Reset the device. All peripherals will be reinitialized.
7.8
Data logger
The optional data logger enables you to record data. Data can be recorded for a period between 6
and 12 months with an interval of 1 second for each parameter. The data can be extracted with
VPStudio 2.0 using the USB connection.
There are 2 different modes available, explained in the 2 paragraphs below. The modes can be
changed with VPStudio 2.0 or by using the key pad.
Changing the mode will remove all stored
data.
Data will only be recorded when a VPSensorCartridge is connected and when the main power is
applied.
7.8.1
Cyclic mode
We developed cyclic data logging mode as an "install and forget" feature. When this mode is
selected. Data will be recorded at all times, as long as power is applied and a sensor is connected.
The data logger will start automatically with a 1 second interval for all available parameters. When the
memory is full, the oldest data will be removed automatically. Total capacity is nearly 6 months.
7.8.2
Multi session mode
This mode enables you to start or stop one session manually by using the key pad. When started,
the data logger will record data on the user defined intervals.
Use the following guidelines for the intervals
Application
Flow
Pressure
Temperature
Standard energy management application
5 min
5 min
5 min
Machine testing - quick fluctuations
1 sec
1 sec
1 sec
Audit - one week
10 sec
10 sec
5 min
Audit - one month
30 sec
30 sec
5 min
Multiple sessions can be recorded. When a session is started, a separate session will be recorded.
It's not possible to append to an existing session.
When a power failure occurs during recording, the session will be stopped. When power is restored,
a new session will start automatically.