VPFlowScope M probe
16
© 2016 Van Putten Instruments BV | MAN-VP-SMPR-UK | Revision:1600 | Date:23-03-2016
Diameter change from
small to large (gradual or
instant)
40 * D
1
5 *D
1
Jet shaped flow
Diameter change from
large to small (gradual
change, between 7 and
15 degrees)
10 * D
1
5 * D
1
Flattened flow profile
1 = inner diameter; 2 = minimum length
6.3
Safety system
Step 1.
1. Remove the compression fitting's parts on the long
threaded end side.
2. Place the safety plate over the compression fitting
long threaded end.
Step 2.
1. Mount the nut, and close tight. Then
mount the compression fitting's teflon
rings and nut.