Vortran GO2VENT 6123 Скачать руководство пользователя страница 3

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 3 

 

II.  Clinical Considerations

 

The GO

2

VENT™ provides short term, pressure cycled, and constant flow 

ventilatory support for either breathing or non-breathing patients. This allows 
the patient to receive consistent and reliable ventilatory support. Because the 
GO

2

VENT™ is pressure cycled, changes in the patient’s lung compliance will 

cause a change in the patient’s breathing rate. The GO

2

VENT™ is positional 

sensitive.  Final adjustments should be made with the GO

2

VENT™ in its 

secured operating position. The GO

2

VENT™ is pressure cycled on inhalation 

and exhalation (PIP and PEEP) which minimizes the possibility of gas trapping. 
During inhalation, exhalation will not start until PIP is reached. During 
exhalation, inhalation will not begin until pressure drops to PEEP. For the 

spontaneously breathing patient, the rate dial of the GO

2

VENT™ is set so the 

baseline pressure is above the intrinsic PEEP allowing the patient to initiate 
inhalation by drawing the baseline pressure down to the set PEEP. Because the 
GO

2

VENT™ is a constant flow pressure cycled device, changes in patient 

compliance will result in changes in the respiratory rate (stiffer or smaller 

compliances produce faster rates). The advantage of this minimizes the danger 
of barotrauma. It should be emphasized that the GO

2

VENT™ is to be used only 

by trained personnel who continuously monitor the patient. The GO

2

VENT™ is 

not an ICU stand-alone ventilator with multiple monitoring features. 

 

Setup and use of the GO

2

VENT™ is simple (refer to Setup Instructions in Section 

III on page 7). Set desired flow (Q), adjust PIP pressure dial to obtain desired 
inspiratory time (t

insp

) to attain tidal volume (TV = Q x t

insp

 see Tidal Volume 

Table 1). The gas flow, patient’s lung compliance, and PIP settings control the 

inspiratory time and tidal volume. Then adjust rate dial to obtain desired 
breathing rate. 

 

Table 1 – Estimated Tidal Volume (mL) Delivered at 

Various Flow (L/min) and Inspiratory Time (Seconds) 

 

Flow 

(L/min) 

Inspiratory Time (Seconds) 

0.5 1 1.5 2 2.5 3 

15 

125 250 375 500 625 750 

20 

167 333 500 667 833 1000 

25 

208 417 625 833 1042 

1250 

30 250 

500 

750 

1000 

1250 

1500 

35 292 

583 

875 

1167 

1458 

1750 

40 

333  667  1000 1333 1667 2000 

 
 

 

 

 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: