Vortran GO2VENT 6123 Скачать руководство пользователя страница 24

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 24 

 

Mario Romano, RCP, Otto G. Raabe, Ph.D, William Walby, MS and Timothy E. Albertson, MD, Ph.D., The Stability 
of Arterial Blood Gases During Transportation of Patients Using the RespirTech PRO™, American Journal of 
Emergency Medicine, May 2000. 

 

PURPOSE

: The transportation of critically ill patients requiring mechanical ventilation is recognized as a high risk and 

expensive procedure. Approaches have included using manual bag-type valve resuscitators and expensive portable transport 
ventilators. This study evaluated the effectiveness of the inexpensive portable RespirTech PRO (RTP) gas-powered automatic 
resuscitator during intra-hospital transport of critically ill mechanically ventilated patients. 

BASIC PROCEDURES

: Twenty medical intensive patients on stable mechanical ventilator settings had arterial blood gas 

and vital sign determination before routine transport out of the intensive care unit. Repeat measurements were made during 
transport approximately 30 minutes after being placed on the RTP portable pressure-cycled automatic resuscitator using an 
FiO2 of 100%. 

MAIN FINDINGS

: During use of the RTP for transport, there were no statistically significant variations observed in mean 

arterial blood pressure [82 ± 11 SD (range 65-112) mm Hg before transport versus 85 ± 14 SD (range 59-110) mm Hg during 
transport], heart rate [94 ± 16 SD (range 74-127) beats/min) before versus 96 ± 17 SD (range 69-132) beats/min during], 
arterial pH [7.41 ± 0.07 SD (range 7.31-7.58) before versus 7.42 ± 0.05 SD (range 7.37-7.52) during], and PaCO2 [43 ± 10 
SD (range 26-65) mm Hg versus during 43 ± 10 SD (range 27-61 mm Hg) during]. Because the FiO2 before transport was 63 
± 26 SD (range 30-100%) versus 100% during transport using the RTP, the mean PaO2 was significantly increased from 124 
± 86 SD (range 57-367) mm Hg before transport to 297 ± 168 SD (range 65-537) mm Hg during (P<0.001). No transportation 
associated clinical adverse events were noted. 

DISCUSSION: 

Several previous investigations have shown that portable ventilators are safe and effective in intra-hospital 

transport of mechanically ventilated patients. This study demonstrated that the portable pressure-cycled RTP also allows safe 
transportation of mechanically ventilated ICU patients. By analogy, the RTP is potentially useful as an automatic resuscitator 
for emergency medical care. 

PRINCIPAL CONCLUSION

: This RTP is a disposable resuscitator/ventilator device that provides an inexpensive alternative 

for transporting ventilator-dependent patients

 

 

Mario Romano, RCP, COMPATIBILITY OF THE RESPIRTECH PRO IN THE MRI UNIT, presented at 46th 
AARC International Respiratory Congress in Cincinnati, OH, October 7-10, 2000.

 

BACKGROUND

: Because most medical facilities do not have MRI compatible ventilators, MRI studies on intubated patients 

are frequently delayed until the patient is extubated. Although there are mechanical ventilators that are MRI compatible, the 
cost for purchasing them for MRI use only is impractical, especially in light of the limited number of intubated patients 
needing an MRI. This paper examines the RespirTech PRO, a single patient use fully automatic resuscitator, and how it 
functioned during an MRI study in a General Electric 1.5 MRI unit. 

METHODS

: One clinically stable 72-year old male patient in need of an MRI of his head was placed on the automatic 

resuscitator with extension kit. The patient was set in a control mode of 16 BPM with a ventilating pressure of 25 cm-H2O 
and a liter flow of 40 LPM at a FiO2 of 100%. The patient was placed in a General Electric 1.5 MRI unit, and the device 
functioned without incident. No attraction to the magnet was noted. Image artifact was minimal and was limited to the patient 
tee area, allowing for a clear picture of the head. The patient tolerated ventilation well, and his vital signs are summarized in 
the graph below. 
 

RESULTS:  

Patient Vital Signs:

 

Tx 

HR 

BP 

O

2

 Sat. 

FiO

2

 Set 

 

 

Pre MRI 

85 

98/51 

98% 

100% 

 

 

During MRI 

77 

102/54 

96% 

100% 

 

DISCUSSION

: No significant changes in vital signs or O

saturation were noted with the use of the automatic resuscitator. 

The patient appeared to tolerate the procedure with no adverse affects. No attraction to the MRI magnet was noted and artifact 
was limited to the patient tee area. 

CONCLUSIONS: 

The RespirTech PRO can be a safe and cost effective ventilator for use in the MRI room without the need 

to purchase capital equipment. More experience with the use of this automatic resuscitator in transporting patients to other 
areas of the hospital can establish it as a safe and cost effective transport.

 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: