Vortran GO2VENT 6123 Скачать руководство пользователя страница 11

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 11 

 

III. Protocol: Setup Instructions - GO

2

VENT™ (continued) 

adjusting rate dial clockwise until mandatory rate stops. To 
return to automatic cycling, rotate rate dial counterclockwise 
until desired respiratory rate is achieved. 

Note: Observe rise and fall of the chest corresponding to patient’s 

inhalation and exhalation. Listen for expiratory flow from 
modulator. Listen to chest sounds.  

Note: If patient vomits, disconnect patient adaptor from modulator 

and remove rate dial if necessary. Tap out vomitus on hard 
surface to dislodge and reassemble. Clear patient’s airway and 
reconnect. Clearing procedure should take less than 20 
seconds. Check that inhalation and exhalation occur without 
obstruction. 

Note: The GO

2

VENT™ is pressure limited and is equipped with a 

redundant pressure pop-off valve which will activate at a 
maximum of 60 cm-H

2

O. 

Note: Changes in patient’s lung compliance will result in respiratory 

rate changes. In such an event, make appropriate clinical 

changes. 

Note: If patient draws air through entrainment port or device is set 

to FiO

2

 of 50%, oxygen concentration delivered to patient may 

differ from concentration at gas inlet of patient connector. 

Note: Perform a FUNCTIONAL CHECK by occluding patient port with 

supply gas flowing and verify that pressure DOES NOT 
EXCEED 60 cm-H

2

O. 

Note: Gas supply source must be capable of delivering up to 40 

L/min.  Typical required supply pressure is 50 ± 5 PSIG. 
Supply pressures from 12 to 80 PSIG may be used if the flow 

is adjusted between 6 to 40 L/min (±10%). 

Note: The  GO

2

VENT™ will deliver 40 L/min against a patient 

pressure of 20 to 40 cm-H

2

O when the green knob is turned 

all the way clockwise and is connected directly to a 50 PSIG 
source. Lower flows are obtainable with flowmeter 
adjustment. 

Note: The GO

2

VENT™ will deliver FiO2 of 50% (±10%) when the 

green knob is turned all the way counterclockwise and is 
supplied with oxygen flow from 6 to 15 L/min with resulting 
output flow of 20 to 40 L/min respectively (see “ENTRAINED 
FLOWCHART”). 

4.4.7  STEP [5]: Adjust Flow, PIP and Rate  

Observe the rise and fall of the chest corresponding to 

inhalation and exhalation of patient. Listen for expiratory flow 
from modulator. Listen to breath sounds of patient. There is 
no substitute for a good clinical assessment. 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: