Vortran GO2VENT 6123 Скачать руководство пользователя страница 23

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 23 

 

Otto G. Raabe, Ph.D. and Mario Romano, RCP, Comparison of RespirTech PRO™ and Ambu

® 

SPUR Resuscitators 

During Simulated CPR. 

BACKGROUND

: The cardiopulmonary resuscitation (CPR) guidelines and American Society for Testing and Materials 

warn against the use of automatic pulmonary resuscitators during CPR closed chest compression because the compression 
process may interfere with lung ventilation and airway resistance may prevent adequate ventilation. However, appropriately 
designed pressure-cycled, pressure-controlled (rather than pressure-cycled, time-controlled) mechanical ventilators should be 
able to automatically respond to pulmonary pressure changes to provide air or oxygen to the lung at high flow rate upon 
demand and alert the rescuer of ventilatory problems. This evaluation was conducted to investigate ventilatory factors 
associated with the use of either the portable RespirTech PRO™ (RTP) gas-powered automatic resuscitator or a typical 
manually operated self-inflating bag-valve resuscitator. 

METHODS

: Thirty tests, 17 with the RTP and 13 with the bag resuscitator, were conducted using the resuscitator connected 

to a commercial test lung modified for automatic simulated chest-compression following standard compression rates as timed 
with an electronic metronome. The test system was designed to be totally mechanical to avoid operator effects. 

RESULTS

: Both resuscitators provided appropriate ventilation without excessive lung pressures following the chosen 5:1 

compression-ventilation ratio. Overall, the RTP (at 25 L/min) and bag-valve resuscitator minute ventilation values were 
about the same with means of 6.3±0.5 SE liters and 6.2±0.6 SE liters, respectively. The RTP automatically responded to 
pulmonary pressure variations, rapidly delivering short breaths between compressions and a full inhalation during the pause 
without serious pressure extremes. The highest observed intrapulmonary pressures (>80 cm H

2

O) occurred with the bag-

valve resuscitator operated during uninterrupted (“seamless”) chest compressions without inhalation synchronization. 

DISCUSSION

: Both devices worked well following the standard protocol for CPR. Because the RTP inhalation-exhalation 

cycling is visually and audibly obvious, indications of possible airway resistance or low tidal volume are readily observed by 
the rescuer. 

CONCLUSIONS

: The RTP may be used safely as an automatic resuscitator during CPR. Revision of CPR guidelines and 

ASTM 920-93 for use of pressure-controlled resuscitators should be considered.

 

 

Michael Rossini, M.D., Barry Hickerson, EMT-P, Preliminary Evaluation of a Lightweight, Disposable Emergency 
Transport Ventilator in the Aeromedical Setting

 

INTRODUCTION: 

Recent evidence suggests patients receiving pre-hospital ventilation benefit from the use of emergency 

transport ventilators (ETV). This evidence is supported by the fact manual ventilation using a bag-valve-mask type device 
has substantial variations in rate and volume. These variations occur during initial treatment and transport even by well-
trained EMS crews. Proper tidal volume, airway pressures and respiratory rate are critical components of emergency 
ventilatory support and variations can impact mortality and morbidity on a wide range of patients suffering from illness or 
injury. 

METHODS

: The purpose of the evaluation was to determine the practicality and ease of use of a new ETV, the “RespirTech 

PRO” manufactured by VORTRAN Medical Technology 1, Inc. and identify any shortcomings during the initial phases of 
patient treatment, transport and emergency room care. The ETV was placed into service on our single BK 117-B2 hospital-
based helicopter program. A Registered Nurse and Licensed Paramedic staff Air Med Team, which is based in Modesto, 
California. The majority of scene transports are flown to our base hospital, Doctors Medical Center also in Modesto, 
California. We gathered data on 12 patients from October 1999 to July 2000 that received ventilatory support from the ETV. 
Vitals signs during and post transport, arterial blood gases post transport and subjective data regarding ease of use, set-up and 
controls where gathered on all 12 patients. 

RESULTS: 

Twelve patients received on-scene and in-flight ventilatory support from the ETV without complications. All 12 

adult patients were intubated by ground EMS personnel or the Air Med Team and placed on the ETV. The two manual 
settings, pressure and rate were set without difficulty and facilitated by the use of continuous end-tidal CO2 monitoring. The 
oxygen source for the ventilator was a 15-25 liter per minute fitting that allowed operation without difficulty in all 12 cases. 
Blood gas analysis and review of vital signs during and post transport indicated all patients had been adequately ventilated 
during initial treatment and transport. 

CONCLUSION: 

The RespirTech PRO proved to be an easy-to-use and reliable ETV that lends itself to a range of patients 

requiring prehospital ventilation. Ventilation is a key factor in the outcome of many types of injury and illness and this ETV 
should be considered for on-scene or transport use in a variety of prehospital settings.

 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: