Vortran GO2VENT 6123 Скачать руководство пользователя страница 27

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 27 

 

XI. Quick Guide

 

 

1

 

Set flow to 10-25 L/min 

Step 1: Set the flow to approximately 25 liters per 
minute with a flowmeter from a suitable gas source. 

Adjust flow as needed.

 

 

 

2

 

Verify PIP ~25 cm-H

2

Step 2: Verify the PIP on the pressure dial. Adjust PIP 
as needed. Each GO

2

VENT™

 

is factory tested and pre-

set to 25 cm-H

2

O.

 

 

 

3

 

Function check - Connect to patient 

Step 3: Before connecting to the patient, perform a 
function check. Connect the GO

2

VENT™

 

to the patient 

and then verify the PIP with a pressure manometer.

 

 

 

4

 

Adjust rate 

Step 4: Adjust the breathing rate with the rate dial. 
Refer to the rate dial label.    

 

 

 

Adjust Flow, PIP and Rate 

Step 5: Re-adjust flow, PIP and rate for the patient’s 
clinical situation. There is no substitute for a good 
clinical assessment.

 

 

 

Use of extension tubing in MRI or CT 

NOTE: Position modulator away from patient tee using 
extension tubing and verify patient cycling whenever 

modulator is repositioned.

 

 

 

This Quick Guide is intended to help you gain a general understanding of the GO

2

VENT™

 

product. Please be certain 

to read, understand, and follow the information listed in this User’s Guide before using this product. 
 

Brief Device Description 

The GO

2

VENT™ provides constant flow, pressure-cycled ventilatory 

support in either pressure control or pressure support modes on 

patients weighing 10kg and above. The device includes the pulmonary 
modulator (an exhalation valve that opens at PIP and closes at PEEP) 
and a patient connector tee to supply gas flow, entrain additional air, 
and provide a redundant pop-off valve for safety. The working 
mechanism of the GO

2

VENT™

 

consists of a moving diaphragm which 

adds or subtracts spring force when it is moved from a horizontal to a 

vertical position, the addition or subtraction of spring force will affect 
the PIP setting by 1~3 cm-H

2

O. The GO

2

VENT™ will function in any 

position as long as the final adjustments are made in a secured 
position 
(strapped or taped to the patient). 

 

PATIENT 

PATIENT 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...