IV. CAUTIONS AND WARNINGS 

CAUTIONS 

Federal law restricts the use of this device to sale by or on order of a physician (or 
properly licensed practitioner). 

WARNINGS 

1.

 

The GO

2

VENT™ should be used only by individuals who have adequate training 

in CPR techniques and in the operation of gas powered resuscitators. 

2.

 

Do not reuse - Risk of cross-contamination. 

3.

 

Do not use grease or oil on the GO

2

VENT™ for any reason. 

4.

 

Spontaneously breathing patients may entrain ambient air. 

5.

 

Supply pressure of 39 to 80 PSIG must be adjustable to 40 L/min. 

6.

 

Redundant pop-off valve is set at 60 cm-H

2

O. 

7.

 

Do not use the GO

2

VENT™ in oxygen deficient atmospheres or near open 

flames. 

8.

 

Do not smoke while using the GO

2

VENT™ or any other oxygen equipment.  

9.

 

Do not dismantle or attempt to remove any components other than those 
required for routine operations. Any tampering with the GO

2

VENT™ may cause 

the unit to malfunction and will automatically void the warranty. 

10.

 

US FDA restricts the use of this device by sale by or on order of a physician (for 

properly licensed practitioner). 

PRECAUTIONS 

1. Patients connected to this device are to be monitored continuously by persons 

having adequate training. Do not leave patients unattended. 

2. When ventilating an intubated patient, higher pressure release settings may be 

required. Select a pressure setting of 35 cm-H

2

O to start and adjust if necessary. 

3. An audible, rapid clicking sound and rapid movement of the diaphragm in the 

modulator indicates airway obstruction. Clear the airway and resume the 
ventilation procedure. 

4. Positive End Expiratory Pressure (PEEP) is intrinsic to this device. PEEP is usually 

1/5

th

 PIP and will range from 2 to 9 cm-H

2

O depending on pressure settings. 

Verify actual PEEP with a manometer. 

5. For a minute ventilation of 10 L/min and an I:E ratio of 1:1: [a] at 100% FiO

2

 

setting - the GO

2

VENT™ will operate for 30 minutes (± 10%) with an output 

and supply flow rate set at 20 L/min on an “E” cylinder volume of 625 liters, [b] 
at 50% FiO

2

 setting - the GO

2

VENT™ will operate for 100 minutes (± 10%) with 

an output flow rate of 20 L/min and supply flow rate set at 6 L/min on an “E” 

cylinder volume of 625 liters.  

6. Please review and follow the instructions and observe the warnings before using 

the GO

2

VENT™. 

7. If the use or operation of the GO

2

VENT™ is unclear, contact your distributor or 

dealer for clarification.  

8. GO

2

VENT™ is a resuscitation management system and should not be used as an 

unattended automatic ventilator. 

 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: