Vortran GO2VENT 6123 Скачать руководство пользователя страница 22

VORTRAN

®

 Medical               

 

5/26/2017 

 
 VORTRAN

®

 GO

2

VENT™ User’s Guide

 

 

Page 22 

 

Steven J. Weiss, Todd Filbrun, Chad Augustin, Ray Jones and Amy Ernst. UC Davis Medical Center: Sacramento, 
CA, Sacramento City Fire/EMS: Sacramento, CA. ABSTRACT: An Automatic Transport Ventilator (ATV) vs. Bag 
Valve Mask (BVM) for Ventilation during EMS Transport. Academic Emergency Medicine Volume 11, Number 5 592, 
May 2004. 

 

ABSTRACT: 

 

OBJECTIVES

: The hypothesis of this study was that paramedics (EMTPs) perceived that use of automatic transport 

ventilator (ATV) was better than BVM for managing ventilation during patient transport.  

METHODS

: ATVs and BVMs were randomized on 5 City Fire Department Paramedic Units. At the conclusion of each 

patient transport, using a 5-point Likert scale, EMTPs rated the modality used (ATV vs. BVM) on ease of use, time of 
setup, expedition of transport, additional tasks completed, documentation, overall patient care, and patient comfort. Pulse, 
oxygen saturation, respiratory rate, and end tidal CO2 were collected every 5 seconds. Statistical analysis was performed on 
results of the Likert scale using a Mann-Whitney U rank sum test. Results were significant if p < 0.05. The power of the 
study was 80 percent to show a difference of 1.0 on the Likert scale.  

RESULTS

: 28 patients were entered into the study, 14 BVM and 14 ATV. The reason for device use was assisted ventilation 

in 7/28 (25%) cases and CPR in 21/28 (75%) cases. There were no significant differences in the EMS perceptions of ease of 
use (p = 0.08), time of setup (p = 0.14), expedition of transport (p = 0.27), or overall patient care (p = 0.59). There were 
significant differences in favor of the ATV in ability to accomplish additional tasks (p = 0.01), ability to document (p = 0.04), 
and ability to provide patient care (p = 0.03). The data collector stored ongoing physiologic data on 15/28 (54%) patients 
during EMS transport. 

CONCLUSIONS

: EMTPs perceived that they were able to accomplish more tasks, document more completely, and 

provide better patient care with the use of the ATV. The data collector time marked data and stored the data for subsequent 
retrieval in a majority of cases.

 

 

Nates, Joseph L. MD, FCCM, Combined external and internal hospital disaster: Impact and response in a Houston 
trauma center intensive care unit*. Critical Care Medicine. 32(3):686-690, March 2004.

 

ABSTRACT: OBJECTIVE: 

To increase awareness of specific risks to healthcare systems during a natural or civil disaster. 

We describe the catastrophic disruption of essential services and the point-by-point response to the crisis in a major medical 
center. 

DESIGN: 

Case report, review of the literature, and discussion.  

SETTING: 

A 28-bed intensive care unit in a level I trauma center in the largest medical center in the world.  

CASE: 

In June 2001, tropical storm Allison caused >3 feet of rainfall and catastrophic flooding in Houston, TX. Memorial 

Hermann Hospital, one of only two level I trauma centers in the community, lost electrical power, communications systems, 
running water, and internal transportation. All essential hospital services were rendered nonfunctional. Life-saving equipment 
such as ventilators, infusion pumps, and monitors became useless. Patients were triaged to other medical facilities based on 
acuity using ground and air ambulances. No patients died as result of the internal disaster.  

CONCLUSION: 

Adequate training, teamwork, communication, coordination with other healthcare professionals, and strong 

leadership are essential during a crisis. Electricity is vital when delivering care in today’s healthcare system, which depends on 
advanced technology. It is imperative that hospitals take the necessary measures to preserve electrical power at all times. 
Hospitals should have battery-operated internal and external communication systems readily available in the event of a 
widespread disaster and communication outage. Critical services such as pharmacy, laboratories, blood bank, and central 
supply rooms should be located at sites more secure than the ground floors, and these services should be prepared for more 
extensive performances. Contingency plans to maintain protected water supplies and available emergency kits with batteries, 
flashlights, two-way radios, and a nonelectronic emergency system for patient identification are also very important. Rapid 
adaptation to unexpected adverse conditions is critical to the successful implementation of any disaster plan.

 

 

 

Содержание GO2VENT 6123

Страница 1: ...onal and Operational Characteristics 2 Figure 1 GO2VENT Component Description 2 II Clinical Considerations 3 Table 1 Estimated Tidal Volume ML Delivered 3 Figure 2 Airway Pressures PIP PEEP 6 III Prot...

Страница 2: ...n port 5 redundant pressure pop off valve and 6 one way valve for entraining additional air The pulmonary modulator provides the actual ventilatory support The primary working mechanism of the pulmona...

Страница 3: ...eline pressure down to the set PEEP Because the GO2VENT is a constant flow pressure cycled device changes in patient compliance will result in changes in the respiratory rate stiffer or smaller compli...

Страница 4: ...cycling the patient s airway may be occluded or a very large leak exists The PIP may be adjusted from 10 and 50 cm H2O The PEEP is intrinsic to the device which ranges from 2 to 9 centimeters and is...

Страница 5: ...ient s airways the GO2VENT will stop cycling or may sometimes cycle rapidly The GO2VENT will work with any mask that provides a good seal with the patient All clinicians should receive adequate traini...

Страница 6: ...red to manual resuscitators is the ability to deliver consistent reliable and hands free resuscitation Manual resuscitators may have adverse effects on patients as a result of inconsistent ventilation...

Страница 7: ...al of the GO2VENT is set so that the baseline pressure is above the set PEEP allowing the patient to initiate inhalation by drawing the baseline pressure down to the set PEEP The device includes the p...

Страница 8: ...ts other than those required for routine operations Any tampering with the GO2VENT may cause the unit to malfunction and will automatically void the warranty 4 4 SET UP INSTRUCTIONS 4 4 1 The GO2VENT...

Страница 9: ...connect the GO2VENT to this port you will automatically get 40 L min Note The GO2VENT is completely gas driven requiring no electrical power and will deliver 100 oxygen to a patient Note The duration...

Страница 10: ...sted to 40 L min 10 The GO2VENT will deliver 40 L min against a patient pressure of 20 to 40 cm H2O when connected directly to a 50 PSIG source Lower flows are obtainable with flowmeter adjustment Use...

Страница 11: ...r through entrainment port or device is set to FiO2 of 50 oxygen concentration delivered to patient may differ from concentration at gas inlet of patient connector Note Perform a FUNCTIONAL CHECK by o...

Страница 12: ...to this device are to be monitored continuously by persons having adequate training Do not leave patients unattended 2 When ventilating an intubated patient higher pressure release settings may be req...

Страница 13: ...athing patient 2 2 To be able to troubleshoot and correct any problem that may arise with the use of the GO2VENT a Gas consumption during use b What is happening if it stops cycling while adjusting th...

Страница 14: ...tion mandatory breathing For patients taking spontaneous breaths requiring assisted breathing If the rate dial has been adjusted to a position that the continuous flow of gas creates more pressure tha...

Страница 15: ...T rate dial controls rate by controlling the exhalation time Once the PIP and inspiratory flow L min have been set inspiratory time is also set The only way to control respiratory rate is by controlli...

Страница 16: ...ed with an orifice type 15 L min flowmeter are also equipped with a high flow power take off port If you connect the GO2VENT to this port you will automatically get 40 L min If 40 L min is too much fl...

Страница 17: ...l treatment while the patient is connected to the GO2VENT Yes NOTE Deposition of medicine residue may cause the GO2VENT to stick if it dries for an extended period of time Always perform a functional...

Страница 18: ...ed resuscitator Bag Valve Mask BVMs are the most commonly used devices for emergency short term ventilator support They are typically disposable and are used extensively in the pre hospital and inter...

Страница 19: ...lator EM 100 is relatively heavy non disposable and is not equipped with a pop off valve The cost is considerably higher than the GO2VENT 30 What are the advantages of the transport ventilators Transp...

Страница 20: ...elines defining a narrow range of optimal oxygen saturation for many situations Based on these recommendations proper patient care mandates that we have the ability to control both components of venti...

Страница 21: ...d with air leaks present and would need prompt medical intervention Although all three units performed as advertised each unit had individual characteristics that would have to be evaluated by the pot...

Страница 22: ...ty of cases Nates Joseph L MD FCCM Combined external and internal hospital disaster Impact and response in a Houston trauma center intensive care unit Critical Care Medicine 32 3 686 690 March 2004 AB...

Страница 23: ...ASTM 920 93 for use of pressure controlled resuscitators should be considered Michael Rossini M D Barry Hickerson EMT P Preliminary Evaluation of a Lightweight Disposable Emergency Transport Ventilato...

Страница 24: ...TP also allows safe transportation of mechanically ventilated ICU patients By analogy the RTP is potentially useful as an automatic resuscitator for emergency medical care PRINCIPAL CONCLUSION This RT...

Страница 25: ...mask intermittent assist device with continuous positive airway pressure device INSTRUCTIONS Coverage issue CIM 60 9 CPT Current Procedure Terminology American Medical Association PRODUCT GO2VENT Auto...

Страница 26: ...Supply gas flow too low increase supply gas flow 2 High PIP setting lower PIP setting as appropriate 3 Compliance too low change mode of ventilation Inspiratory time is too short 1 Supply gas flow too...

Страница 27: ...xtension tubing and verify patient cycling whenever modulator is repositioned This Quick Guide is intended to help you gain a general understanding of the GO2VENT product Please be certain to read und...

Отзывы: