G
ETTING
S
TARTED
IEC61000-3
S
OFTWARE FOR
W
INDOWS
U
SER
M
ANUAL
Page 54
VPN 98-025/7
Voltech Instruments
factor should be within 10% of the user specified fundamental current and
power factor.
Click anywhere on this window to move the red marker and view details
for the associated reading at the top of the main window.
POHC and THC
See the reference information for definitions of these terms.
POHC is used during limit checking (see the reference information) .
Measurement of THC is not a specific requirement of the standard other
than in the absence of specific guidelines, 'emission tests shall be
conducted with the user's operation controls or automatic programs set to
produce the maximum THC under normal operating conditions'.
Click anywhere on this window to move the red marker and view details
for the associated reading at the top of the main window.
Source Qualification
This is a table view of the source voltage, frequency and harmonic content.
Each parameter is allocated a PASS or FAIL according to the requirements
of the standard.