For AT5600 only
Open Circuit Compensation Error:
The external source is not connected to the Open Circuit Fixture.
Internal fault condition: Safety interlock open:
“Safety Interlock Open” message on the AT5600 display as shown below.
Should this error occur; check the operation and installation of the safety light curtain (see AT5600 user
manual).
Voltage Present Condition:
When the AT5600 detects a voltage level >40V preceding a test, the error message shown below is displayed
on the AT5600’s front panel.
Should this error occur,
a. DO NOT TOUCH THE PART UNDER TEST OR THE FIXTURE.
b. Switch off the AC source.
c. Switch off the AT3600.
d. Check to make sure the part under test is safe to touch before removing it and powering on
the AT3600 and AC source.