![VOCARE vitals360 Скачать руководство пользователя страница 113](http://html.mh-extra.com/html/vocare/vitals360/vitals360_user-manual_1043425113.webp)
Instructions for Use
Page
113
of
117
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
150 kHz to 80 MHz
6 V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
150 kHz to 80 MHz
6 V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where
p
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
b
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic is affected by absorption and
reflection from structures, objects and people.
a
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to
6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The
amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz,
21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
P
V
d
]
12
[
2
=
P
E
d
]
5
.
3
[
1
=
P
E
d
]
7
[
1
=
Содержание vitals360
Страница 1: ...Instructions for Use Page 1 of 117 ...
Страница 20: ...Instructions for Use Page 20 of 117 02 DEVICE OVERVIEW 02 1 Know Your Monitor a Device ...
Страница 36: ...Instructions for Use Page 36 of 117 Home Button Back Button returns to the previous screen ...
Страница 37: ...Instructions for Use Page 37 of 117 03 6 Symbols Bank Home Page LANGUAGES ...
Страница 56: ...Instructions for Use Page 56 of 117 04 9 Troubleshooting ...
Страница 61: ...Instructions for Use Page 61 of 117 05 3 Display Symbols Bank Pulse Rate Waveform ...
Страница 72: ...Instructions for Use Page 72 of 117 ...
Страница 78: ...Instructions for Use Page 78 of 117 07 5 Illustrated Example of Measuring ECG 07 6 Troubleshooting ...
Страница 92: ...Instructions for Use Page 92 of 117 ...