2553/ 2553E Command Guide
7.
Check that any device drivers for your controller and operating system are correctly
configured to support the selected 2553/ 2553E address.
I send a device clear but it doesn’t seem to do anything. What is
wrong?
Some instruments perform operations similar to a power cycle when they receive a
device
clear
. This activity is immediately visible on the front panel of the product. The 2553/
2553E fully implements a device clear in less than 350ms, and does not perform a full
power cycle reset. There is little visible activity on the front panel when the IEEE488
bus receives a device clear. This is normal.
The 2553/ 2553E sometimes does not go into REMOTE. What is
wrong?
Some controllers have a configuration to not re-address devices that were previously
addressed. In order for a device to go from LOCAL into REMOTE both the REN line
must be asserted, and the device listen address must be received.
This situation arises if the device enters the LOCAL state when it was previously
addressed to listen, and no other bus address has been used since that time (this could
have been days, or even weeks ago). The controller will not place the 2553/ 2553E (or
other device) back into REMOTE because it will not re-address it, when configured in
the LOCAL state.
In this case, any data apparently sent to the 2553/ 2553E will not actually arrive there and
be acted upon, although the controller acts as if it had been correctly sent. Thus, any
commands sent to the 2553/ 2553E are ignored, as well as the 2553/ 2553E not entering
the REMOTE state.
Some common causes of the 2553/ 2553E entering the LOCAL state are –
•
Executing a device clear bus command
•
Executing a *RST IEEE488.2 command
•
Power cycling the 2553/ 2553E
•
Removing the IEEE488 cable from the 2553/ 2553E and then replacing it
•
Pressing the F4 soft menu key of the front panel of the 2553/ 2553E (marked
REMOTE while in remote). This can be disabled by entering the REMOTE WITH
LOCAL LOCKOUT state.
If these circumstances could occur, then reconfigure the IEEE488 controller to always
readdress a listener.