![Vitrek Xitron 2553 Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/vitrek/xitron-2553/xitron-2553_command-manual_1037508064.webp)
64
2553/ 2553E Command Guide
W-INRUSH[
phase
/DC]
The highest inrush real power using only the DC components. If configured as AC only
then this result is always zero. (See the AC-ONLY command). The result format is
floating point, in Watts.
W-INRUSH[
phase
/RMS]
The highest inrush real power measurement result, including DC if configured (see the
AC-ONLY command), over the full configured bandwidth (see the BANDWIDTH
command), including any non-harmonic content. The result format is floating point, in
Watts.
Defining Frequency Measurement Results
FREQ
The measured frequency of the phase A voltage input unless SYNC=1 command is
prevailing, in which case it is the measured frequency of the current input. The result
format is floating point, in Hz, and is always positive. A zero result indicates that the
input has too low of an AC amplitude for frequency measurement (approx. 5% full-scale
for the option fitted), or that the frequency lies outside of the configured bandwidth. (See
the BANDWIDTH command.)
Defining Integrated Measurement Results
A-HR[
phase
/DC]
The integrated mean amplitude of the current input signal. If configured as AC only then
this result is always zero. (See the AC-ONLY command). The result format is floating
point, in Amp-Hours. TOTAL is invalid for this data.
A-HR[
phase
/RMS]
The integrated RMS current amplitude measurement result, including DC if configured
(see the AC-ONLY command), over the full configured bandwidth (see the
BANDWIDTH command), including any non-harmonic content. The result format is
floating point, in Amp-Hours, and is always positive. TOTAL is invalid for this data.
A-INTEG-AVG[
phase
/RMS]
The average integrated RMS current amplitude measurement result, including DC if
configured (see the AC-ONLY command), over the full configured bandwidth (see the
BANDWIDTH command), including any non-harmonic content. The result format is
floating point, in Amps, and is always positive. TOTAL is invalid for this data.
Содержание Xitron 2553
Страница 1: ...IEEE488 COMMAND GUIDE 2553 2553E Three Phase Power Analyser ...
Страница 2: ......
Страница 10: ...10 2553 2553E Command Guide ...
Страница 38: ...38 2553 2553E Command Guide ...