2553/ 2553E Command Guide
A-RELHARM[
phase
/
harmonic
:
harmonic
]
The RMS current amplitude measurement result evaluated for each harmonic, relative to
the fundamental current amplitude measurement. Each result format is floating point, in
percent, and is always positive. TOTAL data is the mean for all phases.
A-RELHARM[
phase
/
harmonic
-
harmonic
]
The RMS current amplitude measurement result evaluated for a range of harmonics,
relative to the fundamental current amplitude measurement. The result format is floating
point, in percent, and is always positive. This is typically used to obtain the %THD,
relative to fundamental, over a specified range of harmonics. TOTAL data is the mean for
all phases.
EVEN-TRIPLENS[
phase
/
harmonic
-
harmonic
]
The RMS current amplitude measurement result evaluated for a range of harmonics,
using only those harmonics, which are divisible by 3 with an even result (e.g., 6, 12, 18
etc.). The result format is floating point, in Amps, and is always positive. TOTAL is
invalid for this result definition.
K-FACTOR[
phase
/
harmonic
-
harmonic
]
The current K-factor measurement result evaluated for a range of harmonics. The result
format is floating point, unitless, and is always positive. Note that the harmonic range
should always include the fundamental (i.e. 1). TOTAL is invalid for this result
definition.
ODD-TRIPLENS[
phase
/
harmonic
-
harmonic
]
The RMS current amplitude measurement result evaluated for a range of harmonics,
using only those harmonics that are divisible by 3 with an odd result (e.g., 3, 9, 15 etc.).
The result format is floating point, in Amps, and is always positive. TOTAL is invalid for
this result definition.
TRIPLENS[
phase
/
harmonic
-
harmonic
]
The RMS current amplitude measurement result evaluated for a range of harmonics,
using only those harmonics that are divisible by 3. The result format is floating point, in
Amps, and is always positive. TOTAL is invalid for this result definition.