Sending Commands to the 2553/ 2553E
21
Commands That Require No Data
These are individual commands performing a single function that is apparent from the
command itself.
Resetting Commands
*CLS
The action of this command follows the recommendations of the IEEE488.2 standard,
clearing the serial poll response byte and all Transmit Buffer Banks and the definitions of
their contents.
Do not include any commands that define responses from the 2553/ 2553E in the same
set of commands as contains the *CLS.
*RST
The action of this command follows the recommendations of the IEEE488.2 standard,
acting as if the 2553/ 2553E had its power cycled off and back on.
Do not include other commands in the same set of commands as contains the *RST
command.
Follow the *RST command with a delay of at least 350ms prior to other activity on the
IEEE488 bus involving the 2553/ 2553E.
During operation of this command, the 2553/ 2553E will become totally disconnected
from the IEEE488.1 bus for a short period of time. If the 2553/ 2553E is the only non-
controller device on the bus, this may cause the IEEE488.1 controller to issue an error
concerning the lack of IEEE488 devices on the bus. Ignore the error in the situation
described here.
Clearing Measurement Results
CLR-INRUSH
This command clears all inrush measurements, each inrush measurement result is flagged
to be set to the next available applicable measurement.
No other measurement results are affected by this command.
Sending this command while measurement results are frozen has no affect.
This command is implied whenever measurements are started using the MEASURE
command.