Sending Commands to the 2553/ 2553E
29
•
The user transmits: CLR-INTEGRATE;INTEGRATE=START<NL>
The 2553/ 2553E will start updating the integrated results after clearing any
previously integrated values. Note that these commands may be issued in either
order.
MEASURE
This command is used to set whether the measurement results are to be updated or frozen.
The MEASURE command takes a single data only, which may be any of the following
ASCII data.
ASCII Data
Command Action
0
All results are frozen
STOP
All results are frozen
1
Results are cleared and allowed to be updated
START
Results are cleared and allowed to be updated
Table 8 – Data for the MEASURE command
Notes –
•
MEASURE=1 or MEASURE=START always clears any presently stored historical,
captured waveform, and integrated data.
•
MEASURE=1 or MEASURE=START always disables update of integrated data,
and enables update of historical, captured waveform, and standard measurement
data.
•
MEASURE=0 or MEASURE=STOP always disables update of all data, setting all
results as frozen.
•
The MEASURE? Interrogative command returns the present state of the MEASURE
command.
Example –
•
The user transmits: MEASURE=STOP<NL>
The 2553/ 2553E will stop updating all results, leaving all results frozen with the
present contents. If results were already frozen, then this has no affect.
SYNC
This command selects the type of synchronization, which the 2553/ 2553E will use for all
measurement results. The SYNC command takes a single ASCII character as detailed in
the table below.