27
VITA V60 i-Line
®
21.3 Temperature setting
Temperature calibration using the silver test function
Using this program together with the VITA Silver Test Set (VITA order
no. B 230), the temperature in the firing chamber can be verified and
readjusted within a range of plus / minus 40 °C. During calibration, ensure
that the instructions for carrying out the silver test are strictly observed
(see instructions provided with the silver test set). Deviations can result in
incorrect measurements, and as a result, incorrect settings.
To verify the temperature in the firing chamber using the silver test function,
a program must be run at 955 °C and at 963 °C.
Starting the first silver test
:
Press the "1. Silvertest" button, the program is displayed. Place the silver
test inside, press the start button. The program is started and the program
sequence is shown.
Result of the first silver test
:
Silver has thickened slightly, but has not melted
The first silver test was successful. The second silver test can now be carried
out.
Silver has melted
The first silver test was not successful. Enter an estimated correction value
using the + button and repeat the first silver test.
Once the program sequence has been completed, press the settings button in
the program overview followed by the temperature setting button to return to
this menu.
Starting the second silver test
:
Press the "2. Silvertest" button, the program is displayed. Place the silver
test inside, press the start button. The program is started and the program
sequence is shown.
Result of the second silver test
:
The silver has melted to a ball shape
The second silver test was successful. The temperature has now been
correctly set.
The silver has not melted
The second silver test was not successful. Enter an estimated correction
value using the - button and repeat the second silver test.
The second silver test was not successful. Enter an estimated correction
value using the - button and repeat the second silver test.
Result of the silver test
:
Temperature too low
: Enter the correction value using the minus button
Temperature too high
: Enter the correction value using the plus button
Modifications are saved automatically.
1. Silvertest
Start 955 °C
2. Silvertest
Start 963 °C
Содержание V60 i-Line