11
VITA V60 i-Line
®
12 CE marking
With the CE mark, a legally binding declaration is made to the effect that the
device corresponds to the fundamental requirements of directive 2006/95/EC
(Low Voltage Directive) and directive 2004/108/EC/EWG (EMC Directive).
13 Mains power supply failure
The device is equipped with power failure protection. This component
prevents program interruption and therefore any misfiring in the event of a
brief failure of the mains power supply. The power failure protection becomes
effective as soon as the mains power supply fails when a firing program is
running.
Failure of the mains power supply for less than approx. 1 minute
30 seconds:
The program continues to run when the mains power supply returns and is
not interrupted. The display does not operate during this time. The display
shows the current program again when the mains power supply returns.
Failure of the mains power supply for more than approx.1 minute
30 seconds:
The program is interrupted. The display is no longer in operation. When the
mains power supply returns, a signal tone is emitted for 3 seconds and the
display indicates that firing cannot be continued.
When the mains power supply returns, restart takes approx. 1 minute
30 seconds.