Visiomed - LX-360 - 122015
24
25
LX-360
0197
- The screen is showing the message HI: the temperature taken is higher than the range of the mode
selected, either higher than 42.9°C (109.2°F) in
BODY
mode or 60°C (140°F) in
SURFACE TEMP
mode.
- The screen is showing the message LO: the temperature taken is lower than the range of the mode
selected, either lower than 32°C (89.6°F) in
BODY
mode or 0°C (32°F) in
SURFACE TEMP
mode.
This message appears in different circumstances - please find below the most common:
Causes of a Lo Message
Solutions
Reading hampered by hair, sweat…
Make sure there is no obstruction.
Reading hampered by an air flow…
Make sure you are not in a draught; this may interfere
with the infrared system.
The reading was taken too close, the
ThermoFlash®
didn’t have enough time to reset.
Respect a delay of 15 seconds between each reading
- 1 minute is advised.
Measuring distance is too far.
Respect a measuring distance between 5 to 8 cm (2 to
3.14 in), away from the forehead.