ViOptix Intra.Ox 2.0 Скачать руководство пользователя страница 26

 

 

Part number: 

OXY-2-DUR-IFU-1 Rev A

 

 

 

CONFIDENTIAL 

 

 

Page 26 of 26

 

7.6

 

Glossary of Symbols  

 

Sterile Disposable Intra.Ox

TM 

Sheath is 

intended for single-patient one-time use.  

DO NOT REUSE, DO NOT RE-STERILIZE 

 

Sterilized by Ethylene Oxide  

 

Use by expiration date stamped 

 

Attention, consult accompanying 
documents 

 

 

Model Number (Catalog Number) 

 

Serial Number 

 

Manufactured By 

 

Applied part type BF 

 

 

Do not use this equipment in the MRI 
scan room 

 

 
 

Содержание Intra.Ox 2.0

Страница 1: ...andheld Tissue Oximeter Instructions for Use ViOptix Incorporated 39655 Eureka Drive Newark CA 94560 Phone 510 226 5860 Fax 510 226 5864 Website www vioptix com ViOptix part number for this IFU OXY 2...

Страница 2: ...lements 8 4 1 Display 8 4 2 Sensing Surface Sensor Face 9 5 Operating the ViOptix Intra Ox 9 5 1 Device Setup 9 5 2 Holding the Device 13 5 3 Measuring Percent Oxygen Saturation 14 5 4 Other Device Mo...

Страница 3: ...ld Tissue Oximeter is constructed from biocompatible materials that can tolerate bodily fluids and other liquids such as disinfectants and marking materials This manual has been prepared to assist med...

Страница 4: ...mation in this manual has been carefully checked and is believed to be accurate In the interest of continued product development ViOptix reserves the right to make changes and improvements to this man...

Страница 5: ...intended to be used in an MR environment Do not look directly at the light emitting distal tip of the Intra Ox Inspect the sensor before each use for visible damage Do not use the instrument if the se...

Страница 6: ...nt Consult the manufacturer for help 2 1 3 Precautions Precautions alert the operator to conditions that could lead to tissue irritation or erroneous results Clean tissue if colored disinfectants are...

Страница 7: ...ed of with electrically hazardous medical waste and in accordance with all local and hospital disposal procedures 2 1 4 3 Battery Pack The Intra Ox Handheld Tissue Oximeter battery pack contains lithi...

Страница 8: ...ternal power source Before using inspect the device and packaging for any sign of damage Do not use if unit has been compromised No installation is required 4 Device Configuration and Interface Elemen...

Страница 9: ...5 1 Device Setup The procedure for Intra Ox setup requires two operators A sterile operator inside the sterile field This operator handles the disposable kit including the sterile sheath A non steril...

Страница 10: ...perator obtains the double bagged sheath and double bagged optical QC target from disposable kit The outer pouches are non sterile and the inner pouches are sterile 2 The outer pouches should be opene...

Страница 11: ...Right The ViOptix logo After a few seconds the device will authenticate the battery and display the following screen Figure 4 Battery and configuration authentication screens After battery authentica...

Страница 12: ...ted their part of the installation process Once the sheath is latched closed it should not be opened again for the duration of the procedure After installation the system will verify the sheath instal...

Страница 13: ...terile operator will hold the distal tip of the assembled device in firm parallel contact against the plastic After this check the device transitions into measurement mode and the Intra Ox is ready fo...

Страница 14: ...play is organized as depicted below Figure 5 Normal Display Layout 5 3 1 Powered Up Time Display The total elapsed time that the device has been powered up is displayed in hours minutes and seconds in...

Страница 15: ...ith the tissue If at least a 2 cannot be achieved try taking a measurement on an adjacent measurement site 5 3 4 Percent Oxygen Saturation Display The estimated oxygen saturation is indicated in perce...

Страница 16: ...Low Battery Indicator The Intra Ox tissue oximeter can continue to be used as long as battery life remains The battery is not rechargeable 5 4 Other Device Modes 5 4 1 Standby Mode If there are no va...

Страница 17: ...ield in the operating room In this mode the device will display the following guidance It will then ask the user to re do the optical QC target checks outlined in Section 0 Assuming that the optics ar...

Страница 18: ...ent requires preparation to get ready for the next procedure This requires a two step process 5 5 2 1 Cleaning the sensor face Immediately after the procedure the sensor face at the distal tip of the...

Страница 19: ...an be taken following the display of one of these warnings OLED Message Recommended User Action OVERHEATED Place down to cool Move the device to a cooler area until condition no longer exists Too much...

Страница 20: ...ure Replace Durable The device is inoperable Battery Depleted Replace Battery The main unit must be paired with a new battery back and sheath to continue measurements Sheath Opened The sheath was open...

Страница 21: ...normally in a humidity environment of 20 to 80 non condensing Transport and Storage Device packaged shall operate normally after storage at 18 C to 60 C Device packaged shall operate normally after st...

Страница 22: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 22 of 26 7 Labels 7 1 Box and Pouch label Sterile Sheath...

Страница 23: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 23 of 26 7 2 Box label Re Usable Main unit...

Страница 24: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 24 of 26 7 3 Box and label Disposable Battery...

Страница 25: ...r OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 25 of 26 7 4 Box and Pouch label Sterile Optical QC Target 7 5 Device label Placed on the back of the re usable main unit Placed on the front of the re usable...

Страница 26: ...M Sheath is intended for single patient one time use DO NOT REUSE DO NOT RE STERILIZE Sterilized by Ethylene Oxide Use by expiration date stamped Attention consult accompanying documents Model Number...

Отзывы: