![ViOptix Intra.Ox 2.0 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/vioptix/intra-ox-2-0/intra-ox-2-0_instructions-for-use-manual_1028931018.webp)
Part number:
OXY-2-DUR-IFU-1 Rev A
CONFIDENTIAL
Page 18 of 26
5.5
After the procedure
5.5.1
Disassembly
Once the sterile field is broken, the Intra.Ox™ device may be removed from the sheath by
pushing upwards on the metal latch on the front of the device. This will result in the ‘Sheath
Open Mode’ warning appearing on
-screen. The re-usable main unit and battery pack can be
pulled out of the sheath.
To power off the device, pull apart the battery pack and the re-usable main unit. (see figure
below)
The battery pack and sheath should be disposed-of in accordance with Section 2.1.4.3.
5.5.2
Preparing the re-usable main unit component for the next procedure
After removing the battery pack, the re-usable main unit component requires preparation to get
ready for the next procedure. This requires a two-step process
5.5.2.1
Cleaning the sensor face
Immediately after the procedure, the sensor face at the distal tip of the device should be wiped
down with a wet cloth or gauze pad to remove any traces of adhesive from the window. These
may interfere with device performance if not removed from the device prior to the next device
use.
5.5.2.2
Cleaning the bulk device
After the procedure, the re-usable main unit component should be cleaned and disinfected using
CaviWipes
TM
. First, use a CaviWipe
TM
to clean the device, removing any visible debris from the
surfaces of the re-usable main unit component. Then, using additional CaviWipe
TM
towellettes
disinfect all device surfaces by ensuring all surfacesare wet for at least three minutes. After the
3 minutes of disinfection time allow the device to dry for at least 10 minutes prior to use.
Содержание Intra.Ox 2.0
Страница 22: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 22 of 26 7 Labels 7 1 Box and Pouch label Sterile Sheath...
Страница 23: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 23 of 26 7 2 Box label Re Usable Main unit...
Страница 24: ...Part number OXY 2 DUR IFU 1 Rev A CONFIDENTIAL Page 24 of 26 7 3 Box and label Disposable Battery...