![ViOptix Intra.Ox 2.0 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/vioptix/intra-ox-2-0/intra-ox-2-0_instructions-for-use-manual_1028931021.webp)
Part number:
OXY-2-DUR-IFU-1 Rev A
CONFIDENTIAL
Page 21 of 26
6
Specifications
Component
Specification
Tissue Oxygen
Saturation
Range
1
–
99%
Wavelengths
730, 760, 810, 845, 895 nm LED
System Control
System will perform self
‐
test when power is turned on
Alarms (visual)
Display conditions such as
•
Low battery
•
Measurement errors
Battery Life
8 hours of continuous use
Operating
Mode
Spot Measurement Mode
Operating
Conditions
•
Device shall operate normally at ambient temperatures of 18°C to
24°C
•
Device shall operate normally in a humidity environment of 20% to
80% (non-condensing)
Transport and
Storage
•
Device (packaged) shall operate normally after storage at -18°C to
60°C
•
Device (packaged) shall operate normally after storage at 10% to
85% (non-condensing).
Dimensions
3
” x
3
” x
7
”
Weight
Less than 1 lb.
Sensor
Specifications
IEC 60601-1. Any surface of the system that comes into contact with a
patient for a time of 1 minute shall not exceed 41ºC.
The device complies with ANSI/IESNA RP-27.1-15 (Recommended
Practice for Photobiological Safety for Lamps and Lamp Systems
–
General Requirements) for light output.
IPX Rating
•
Main unit Only (No Battery Pack): IPX2
•
Sealed Sheath (with Main unit and battery pack installed): IPX2
•
Battery Pack: IPX0