Page 13 of 26
5.1.7
Optical Target Quality-Control Check
To verify the sheath installation process, the sterile operator uses the enclosed optical QC target.
10.
First, the distal tip of the device is firmly pressed against the black hemispherical side of
the optical QC target. This ensures good contact between the optics and the window.
11.
Next, the device prompts the operator to press the device against the other side of the
optical QC target.
The sterile operator will hold the distal tip of the assembled device in firm, parallel contact
against the plastic. After this check, the device transitions into measurement mode, and the
Intra.Ox™ is ready for use.
5.2
Holding the Device
Hold the tip of the
Intra.Ox™ device similarly to a pen, with the wrist in a neutral position. Lightly
rest hand on the patient and use fingers to brace the device for maximum stability. Sensing face
should be parallel to tissue and in gentle contact. Do not apply pressure to the sample site as
blanching may occur, resulting in local ischemia and a misleading measurement.
Avoid non-parallel/angled contact with the tissue as doing so reduces device measurement
accuracy.