USER'S MANUAL
General use instructions
ENDOS AC/ACP - CE
(Rev. 7)
36
“TOOTH ANATOMIC” selectio
“TOOTH ANATOMIC” selectio
“TOOTH ANATOMIC” selectio
“TOOTH ANATOMIC” selection key
n key
n key
n key
Press key (
8
) to select from the exposure times provided for the
different teeth. The meaning of symbols is explained as follows.
Bite-wing
Upper molar
Lower molar
Premolars
superior or
inferior
Incisors /
Canines
superior or
inferior
Exposure enable key (Ready)
Exposure enable key (Ready)
Exposure enable key (Ready)
Exposure enable key (Ready)
For additional safety, ENDOS ACP timer is provided with an "Enabling of
exposure" function.
X-rays can be emitted only after pressing the key . The
equipment enabled signal is displayed on the keyboard by switching on
the corresponding green signal. This enabled signal stays for a set time
(variable during setup between 10 and 30 seconds: default 15 seconds),
after which the function itself is disabled and X-rays are not emitted if
the exposure button is touched.
NOTE:
When the READY key is pressed not only will exposure be enabled, but
the equipment will also start displaying exposure time on the display
calculated according to line voltage fluctuation compensation (see
paragraph 5.1).
Содержание ENDOS AC
Страница 1: ...Release 30 August 2013 Rev 7 ENDOS AC ACP 0051 User s Manual User s Manual User s Manual User s Manual...
Страница 3: ...USER S MANUAL Revision history ENDOS AC ACP CE Rev 7 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 32: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 27 Curve anode cooling Curve tubehead cooling...
Страница 33: ...USER S MANUAL Technical data ENDOS AC ACP CE Rev 7 28 KL16 0 8 70G Feature of emission Filament characteristics...
Страница 34: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 29 Heat storage kJ Curve tubehead cooling...
Страница 75: ...USER S MANUAL Maintenance ENDOS AC ACP CE Rev 7 70 THIS PAGE IS INTENTIONALLY LEFT BLANK...