![Villa Sistemi Medicali ENDOS AC Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/villa-sistemi-medicali/endos-ac/endos-ac_user-manual_1026491065.webp)
USER'S MANUAL
Messages on display
ENDOS AC/ACP - CE
(Rev. 7)
60
•
X
X
X
X----ray button released before exposure end
ray button released before exposure end
ray button released before exposure end
ray button released before exposure end
The X-ray emission button is a "dead man" control; so it must be
held pressed during the whole exposure. If the button is released
before the exposure ends, the emission is automatically stopped; this
situation is shown on the display by the flashing message "E P", and
the luminous signals on the keyboard will be off. To reset this
condition, press "increase" or "decrease": the green LED will start
blinking without exposure and the tube cooling time countdown will
be shown. When the cool down is over the device goes back to the
IDLE state.
NOTE:
Before starting a new exposure, replace the film to avoid a double
exposure with meaningless diagnostic results.
Содержание ENDOS AC
Страница 1: ...Release 30 August 2013 Rev 7 ENDOS AC ACP 0051 User s Manual User s Manual User s Manual User s Manual...
Страница 3: ...USER S MANUAL Revision history ENDOS AC ACP CE Rev 7 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 32: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 27 Curve anode cooling Curve tubehead cooling...
Страница 33: ...USER S MANUAL Technical data ENDOS AC ACP CE Rev 7 28 KL16 0 8 70G Feature of emission Filament characteristics...
Страница 34: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 29 Heat storage kJ Curve tubehead cooling...
Страница 75: ...USER S MANUAL Maintenance ENDOS AC ACP CE Rev 7 70 THIS PAGE IS INTENTIONALLY LEFT BLANK...