USER'S MANUAL
Technical data
ENDOS AC/ACP - CE
(Rev. 7)
18
5.
5.
5.
5.
TECHNICAL DATA
TECHNICAL DATA
TECHNICAL DATA
TECHNICAL DATA
Technical features
Equipment
ENDOS AC/ACP
Manufacturer
VILLA SISTEMI MEDICALI
Buccinasco (MI)
Class
Class I° with type B applied parts
(EN 60601-1 classification)
Protection level
Standard Apparatus IP20
Line voltage
230 V
∼
±
10%
Line frequency
50 Hz
Absorbed current
4 A rms impulsive @ 230 V
∼
Power consumption
920 VA impulsive @ 230 V
∼
Max. apparent line resistance
0.8
Ω
max.
Main fuse
6 AF
Pre-set exposure times
from 0.02 to 3.2 s in 33 steps
Automatic selection
(only for ENDOS ACP)
30 pre-set times
Exposure time accuracy
±
10% or
±
20 ms
(whichever is greater - see note
paragraph 5.5)
Circuit type
Single phase self-rectifying with grid
control
kV selection (high voltage value)
70 kVp
Tubehead current
8 mA
KV accuracy
±
6 % @ nominal voltage
Tubehead (anode) current accuracy
±
13 % @ nominal voltage
Max. exposure time
3.2 s
Timer dimension
345
×
195
×
100 mm
Содержание ENDOS AC
Страница 1: ...Release 30 August 2013 Rev 7 ENDOS AC ACP 0051 User s Manual User s Manual User s Manual User s Manual...
Страница 3: ...USER S MANUAL Revision history ENDOS AC ACP CE Rev 7 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 32: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 27 Curve anode cooling Curve tubehead cooling...
Страница 33: ...USER S MANUAL Technical data ENDOS AC ACP CE Rev 7 28 KL16 0 8 70G Feature of emission Filament characteristics...
Страница 34: ...USER S MANUAL Technical data Rev 7 ENDOS AC ACP CE 29 Heat storage kJ Curve tubehead cooling...
Страница 75: ...USER S MANUAL Maintenance ENDOS AC ACP CE Rev 7 70 THIS PAGE IS INTENTIONALLY LEFT BLANK...