31
9.3 GLP-compliant print
At calibration, the AJH-CE Series enables printing in a format compliant with GLP. Set the function to
the GLP print support ([
]) side, then connect the balance to a printer or other equipment.
When calibration is complete, the balance prints out the format illustrated below.
Since only items are printed, the data should be entered manually by the person who conducted
calibration.
※
Reference: “4.2 Description of Functions” (Page 13)
“9.2 Calibration of AJH-CE Series” (Page 29)
GLP-Compliant Print Form
**CALIBRATION**
MODEL:
S/N:
ID:
DATA:
TIME:
*CAL.END
NAME:
***************
←見出し
←はかりの型名
←はかりの製造番号
←ID番号
←校正した日付
←校正した時刻
←校正終了
←担当者署名
Title
Model of balance
Serial No. of balance
ID No.
Date of calibration
Time of calibration
End of calibration
Signature of person in charge
**CALIBRATION**
MODEL:
S/N:
ID:
DATE:
TIME:
*CAL.END
NAME:
***************
※
To operate CSP-160, set On in switch 3 (Printing controlled by scale), and OFF in
switch 7(disable Manual printing) in dipswitch setting.
For more information on the printer, refer to the printer’s operation manual.
Caution:
Printing is done only when calibration ends normally. The balance will not print if calibration ends
abnormally with [
], [
], or [
] being issued.