28
4
Calibrating the capacity point
Load the calibration weight in the center
of the balance.
The display flashes, indicating that
capacity point calibration is in progress.
5
Terminating the span adjustment
When the calibration of the capacity
point is finished, the original
measurement mode is restored.
☆
Key Points of the Procedure
☆
1.
Pressing the Function key in Step 2 interrupts the span adjustment and returns you to the
original measurement mode.
2.
The calibration weight used for span adjustment should be heavier than 50% of the capacity of
the balance.
To implement a calibration as precisely as possible, use a weight close to the capacity of the
balance.
※
Calibration weights can be ordered from Shinko. For ordering information, please contact
Shinko.
3.
If problems arise during span adjustments, one or more of the following error messages will
appear:
(1) [
]: The calibration weight exceeds the capacity of the balance.
(2) [
]: The calibration weight is less than 50% the capacity of the balance.
(3) [
]: The difference between before and after calibration values is too large (1.0%
or more).
※
If error messages are displayed, calibration cannot take place.
Check the weight and re-calibrate. If the same error continues after repeated calibrations using
the correct weight, please contact our Marketing Division or Technical Service Division.