9.2 Calibration of AJH-CE Series
Applicable models: AJH-2200CE, AJH-3200CE and AJH-4200CE
※
Span adjustment should be performed with the balance installed perfectly level and without any
load on the measurement pan.
1
Invoke the span-adjustment function.
押し続ける
指を離す
Hold down.
Release your
finger.
Hold down the Function key, then
release your finger when the display
switches from [
] to
[
].
2
Start span adjustment.
Zero/Tareキーを押しながら
Functionキーを押す
② ①
同時に離す
CAL
Press the Function key while
holding down the Zero/Tare key.
Release both
fingers at the
same time.
Press the Function key while holding
down the Zero/Tare key, then
release both fingers simultaneously.
Display will change to a flashing [
] and the zero-point will be calibrated
automatically.
3 Zero-point
calibration
CAL
CAL
Once zero-point calibration is
completed, display switches to [
]. Calibration now proceeds to
that of the weighing capacity point.
4
Calibration of weighing capacity point
CAL
CAL
Calibration
Slowly and fully turn the calibration
knob in the direction of [Calibration].
The balance assumes the state where
the calibration weight is loaded on the
measurement pan.
The display starts flashing and the
balance automatically adjusts the
weighing capacity point.