Chapter 4
Power meter and Source functions on the OTDR test port
Activating the Source function
80
User Manual
700MAN201 Rev. 002
•
Reference level
Enter a reference value for the selected wavelength. Select a
wavelength, and enter the value with the numeric keyboard.
The reference value can be automatically measured, in the
Results
page, using the
Standard Ref.
key.
•
Attenuator compensation
Set the attenuation level per wavelength of the external
attenuator used for the measurement. Thus the power level
displayed is the measured power level plus the compensation
level. Select a wavelength, and enter the value with the
numeric keyboard
.
Configuring the alarm parameters
•
Alarms
Activation of pass/fail analysis: any result below the lower
threshold or above the upper threshold will be displayed in red
on the Results page.
•
Min and max thresholds
Choice of lower and upper thresholds for each available
wavelength, from -60 to +40 dBm.
Activating the Source function
To activate the function:
1
Press the
H
OME
button.
2
Select the light source icon in OTDR functions section.
It turns yellow to show it is active
.
LTS Results page
The LTS results page is split into two sections:
•
the upper section corresponds to the light source activation and settings
CAUTION
The light source icons of the singlemode port and the one of the multimode
port cannot be selected at the same time. When one is selected, it automat-
ically deselects the other one.
OTDR New IHM rev002.book Page 80 Jeudi, 5. mai 2022 11:50 11
Содержание OTDR 41 A Series
Страница 2: ...OTDR New IHM rev002 book Page ii Jeudi 5 mai 2022 11 50 11 ...
Страница 4: ...OTDR New IHM rev002 book Page iv Jeudi 5 mai 2022 11 50 11 ...
Страница 14: ...Contents xiv User Manual 700MAN201 Rev 002 OTDR New IHM rev002 book Page xiv Jeudi 5 mai 2022 11 50 11 ...
Страница 147: ...OTDR New IHM rev002 book Page 129 Jeudi 5 mai 2022 11 50 11 ...