Chapter 3
Launching a reflectometry test and displaying results
Advanced functions in Expert OTDR mode
User Manual
700MAN201 Rev. 002
63
1
Place the cursor A at the beginning of the section of the trace where the slope is
to be measured.
2
Place the cursor B at the end of this section.
3
Press the
Analysis
key
4
Press
Slope
in the manual measurements bar.
5
Press
Calculate
key.
The slope of the specified trace section is displayed.
The result is displayed on the screen between the two slope indicators [ and ].
The measurement results are also available in the table:
Click again on
Anaysis
to return to results table display.
"Distance" shows the distance between the beginning of the trace and the end of the
slope;
If no result is displayed in the table, this means that the distance between the cursors A
and B is too small.
Figure 38
Manual Measurement results
Slope
OTDR New IHM rev002.book Page 63 Jeudi, 5. mai 2022 11:50 11