Chapter 6 Care and Maintenance
Module Removal/Installation Procedures
Rev. C0
mA-3011 AXIe Embedded Host Module Operation Guide
April 2020
Page 6-7
Module Installation Procedure
The mA-3011 is designed for installation in an AXIe Chassis compliant to AXIe-1 Base
Architecture Specification, Revision 3.0.
Preliminary Procedures
Verify the following before beginning installation:
•
Slot does not contain foreign objects or debris.
•
Chassis Backplane connector pins are not bent or damaged.
CAUTION
Modules are not "hot -swappable." The Chassis must be powered down before installing
or removing modules from the Chassis.
Mise en Garde
Les modules ne peuvent pas être « changés lorsque sous tension. » Le châssis doit être
mis hors tension avant d’installer ou d’enlever des modules du châssis.
CAUTION
Use care when installing modules to avoid damaging any modules already installed in the
Chassis.
Mise en Garde
Faites attention lors de l’installation de modules afin d’éviter d’endommager les modules
déjà installés dans le châssis.
CAUTION
Modules are ESD sensitive and should only be installed, removed and/or serviced by
Qualified Service Personnel.
Mise en Garde
Les modules sont sensibles aux DES et ils doivent seulement être installés, enlevés ou
entretenus par du personnel de service qualifié.