![VIASYS Vmax Encore Скачать руководство пользователя страница 80](http://html.mh-extra.com/html/viasys/vmax-encore/vmax-encore_technical-reference-manual_1011585080.webp)
Chapter 3 Maintenance and Troubleshooting
3–26
L2707
Technical Reference Manual
Vmax Encore
Dir
Displays a voltage representing a pressure signal. The sign (positive/negative) of the
zeroed pressure signal represents the direction of the flow
E1
Voltage representing the flow sensor. E1 of 3.2ish represents a valid calibration when
no flow is present (Spectra or Encore)
Pm
Mouth Pressure transducer
Pb
Box Pressure transducer
Temp
Temperature. 2.266 represents 22.66 degrees Celsius
Pbar
Barometric Pressure. 7.516 represents 751.6 mmHg
Ecold
Created during development of Spectra— not currently used for troubleshooting
Ehot
Created during development of Spectra—not currently used for troubleshooting
Eref
Known voltage that is used to scale the A/D converter output.
Ezero
Known zero volt signal used to zero the A/D converter
5/Rs
Displays the status of the F1 remote start button. 5 volts represents button not pushed
or activated. 0 volts represents activation
+15
Used to detect Vmax. If more than 14 volts isn’t present, then problem with connection
of Vmax
Vol
Used to analyze 922/1022 volume
O2
Oxygen sensor output
CO2
Carbon Dioxide sensor output
CH4
Methane sensor output
C2H2
Acetylene sensor output
CO
Carbon Monoxide sensor output
ECG1-2
Used to read the raw voltages from the SMC 3 lead
POD 1-8
Used to read the raw voltages from the analog pod box per channels 1-8
F1 Shunt
When checkmark present the pressure transducers reference the ambient conditions.
When unchecked, both sides of the pressure transducers reference the mass flow
sensors (used to zero the transducer)
F2 BXB/Extrn
When checked gas to the gas sensors comes from the Si/So line. When unchecked,
gas comes from the BxB line.
F3 TTL A
Outputs a TTL logic level of on or off ( 1 or 0)
F4 TTL B
Outputs a TTL logic level of on or off ( 1 or 0)
F5 TTL C
Outputs a TTL logic level of on or off ( 1 or 0)
Содержание Vmax Encore
Страница 1: ...L2707 Rev A 2005 VIASYS Respiratory Care Inc Technical Reference Manual Vmax Encore ...
Страница 4: ...Vmax Encore Technical Reference Manual iv L2707 ...
Страница 6: ...Vmax Encore Technical Reference Manual vi L2707 ...
Страница 26: ...Vmax Encore Technical Reference Manual 1 20 L2707 ...
Страница 54: ...Vmax Encore Technical Reference Manual 2 28 L2707 ...