Chapter 2 System Description
2–18
L2707
Technical Reference Manual
Vmax Encore
Figure 2-11 Output Voltage Gain Comparison to Ideal
The digital equivalent of the Mass Flow Sensor output signal is multiplied by a number determined
during the calibration procedure to correct the gain.
Figure 2-12 Output Voltage Gain Correction
Calibration
This chapter covers the calibration and verification of flow volume and calibration of
plethysmograph pressure.
Note:
Calibrate the system at least once every testing day to ensure accurate test results.
Calibration Setup (Vmax and Autobox)
1.
Select
1 Flow Sensor Calibration
on the Vmax Program Manager screen to access the
Flow Volume Calibration screen.
Note:
For information on all the options accessible from the Flow Volume Calibration
screen, refer to the reference manual.