![VIASYS Vmax Encore Скачать руководство пользователя страница 77](http://html.mh-extra.com/html/viasys/vmax-encore/vmax-encore_technical-reference-manual_1011585077.webp)
Technical Reference Manual
Vmax Encore
L2707
3–23
Chapter 3 Maintenance and Troubleshooting
TROUBLESHOOTING
Troubleshooting a particular problem entails observing the symptoms and eliminating possible
causes until the problem is found.
The best course of action is to first verify that only one device has a problem. If more than one
device has a problem, check if those devices share common processors, voltages, A/D converters
etc.
To troubleshoot a single device, check that device’s signal path from start to finish.
ALL PRESSURE TRANSDUCERS
If all pressure transducers are not responding, check the signals that they all have in common:
+15.0 V dc
–15.0 V dc
+10.0 V dc
–10.0 V dc
BAROMETRIC PRESSURE TRANSDUCER
The barometric pressure transducer is mounted on the Mass Flow Sensor board. Four op-amp
circuits that make up IC U13 condition the transducer signal.
To troubleshoot the barometric pressure transducer, check the op-amp outputs at U13 pins 1, 7, 8,
and 14.
MOUTH PRESSURE TRANSDUCER
The mouth pressure transducer is mounted on the pressure board. The transducer signal is
conditioned by one op-amp circuit part of IC U14 pins 12, 13, and 14.
To troubleshoot the mouth pressure transducer, check the op-amp outputs at U14 pin 14.
DIRECTION PRESSURE TRANSDUCER
The direction pressure transducer is mounted on the pressure board. The transducer signal is
conditioned by one op-amp circuit part of IC U14 pins 2, 3, and 1.
To troubleshoot the direction pressure transducer, check the op-amp outputs at U13 pins 1, 7, 8,
and 14.
Содержание Vmax Encore
Страница 1: ...L2707 Rev A 2005 VIASYS Respiratory Care Inc Technical Reference Manual Vmax Encore ...
Страница 4: ...Vmax Encore Technical Reference Manual iv L2707 ...
Страница 6: ...Vmax Encore Technical Reference Manual vi L2707 ...
Страница 26: ...Vmax Encore Technical Reference Manual 1 20 L2707 ...
Страница 54: ...Vmax Encore Technical Reference Manual 2 28 L2707 ...