Chapter 1 Introduction
1–14
L2707
Technical Reference Manual
Vmax Encore
BAROMETRIC PRESSURE SENSOR
A barometric pressure sensor is located in the Analyzer Assembly to continuously measure
ambient barometric pressure, making operator-entry of a reading from a barometer unnecessary.
This same sensor also measures changes in gas sample line pressure so that adjustments in gas
partial pressures can be calculated.
DIRECTION PRESSURE SENSOR
A direction pressure sensor is located in the Analyzer Assembly to determine the direction of the
flow signal. This sensor provides for automatic detection of inspiratory and expiratory volumes.
MOUTH PRESSURE SENSOR
A mouth pressure sensor is located in the Analyzer Assembly to measure respiratory pressures at
the mouth.
ANALOG I/O POD
An Analog I/O Pod is provided for interfacing external devices. The pod provides for the following
device interfaces:
•
Eight analog input
•
Four analog output
•
Three TTL output
•
One analog heart rate input or SMC 3-Lead ECG input