Battery charger / inverter COMBI-GAMMA
68
090138.02
E7-02: Aux-Relay 1 Invert Select
Setting
Function
E7-02=0
(Initial setting)
Auxiliary Relay 1 function is normal.
E7-02=1
Auxiliary Relay 1 function is inverted.
This is used to invert the Aux-Relay 1 So ON becomes OFF and OFF becomes ON. In
•
the program, the labels are adapted to reflect this inversion.
E7-03: Aux1 not Switch Off Time (Aux-Relay 1 do not switch off with certain period)
Use constant E7-03 to determine the minimum ON time.
•
The Aux-Relay 1 will not be switched off within the time specified here measured from
•
the moment that all on condition are inactive.
Note: OFF conditions with a delay of 0 minute, ignore this setting.
*
E8 Group (Aux-Relay 2 Option)
E9 Group (Aux-Relay 3 Option)
The functions and the settings of E8 and E9 Groups are exactly the same as those of
*
E7 Group so please refer to above E7 Group description and setting for E8 and E9
Groups.
O Group (Operator):
O1 Group (Monitor Select)
O1-01: Power ON LCD Monitor Select
After start up of the Combi-Gamma, the monitor selections will be visible on the LCD
•
Display, U1-05 Battery Voltage is the initial value shown in the display.
All the constants in U1 Group can be programmed (U1-01~U-26).
•
O1-02: Key Idle Detect Time
Use constant O1-02 to set the idle time when the keyboard is not operated and once any key
•
is pressed, the display will return to the LCD monitor selection value set in constant O1-01.
Initial Setting=180 sec, setting range: 10~600 sec.
•
O2 Group (Key Selections)
O2-01: Key Sounds
Setting
Function